DP0004a

DISPLAYS ON THE EVALUATION UNIT

Yellow LEDs:

Selected measuring range: 1 LED = LOW / 2 LEDs = MEDIUM / 3 LEDs = HIGH

Blue LED:

LED lit: Limit value exceeded; LED off: measurement below the limit

0-100:

Selected limit value (%) in the specified measuring range

TR:

Receiver malfunction or receiver not connected

TT:

Emitter malfunction or emitter not connected

CR:

Memory error (perform reset to factory settings)

ST:

Default values (after reset to factory settings)

ER:

Excessive number of transmission errors during digital transmission

HT:

Permissible temperature range of the transducer exceeded

LT:

Valid temperature of the transducer underrun

In the case of Measuring cells for piping systems (TURBISWITCH GA…), the green LEDs in the terminal housing of the TT-HDR emitter and
the TR-HDR receiver blink as long as internal measured value processing operates correctly (a continuously illuminated or red LED indicates a

fault).

LIMIT VALUE SWITCHING PERFORMANCE

The blue LED is off: turbidity of the medium is below the selected limit value. The relay for ’’Below limit value’’ is open.
The blue LED is lit: the limit value has been exceeded. The relay for “limit value exceeded” is open.

LIMIT VALUE ADJUSTMENT

Setting range: 3x 0...100% (LOW / MEDIUM / HIGH) by turning the rotary/pushbutton switch.

Note: Switching to the next higher or lower measuring range is accomplished by simply turning the rotary/pushbutton selector.

The selected hysteresis value is automatically skipped in this case. 100% in the LOW range is followed by 0% in the MEDIUM range.
100% in the MEDIUM range is followed by 0% in the HIGH range.

100% in the HIGH range is the highest possible limit value.

0% Low/Med/High

100% Low

100% Med

100% High

Overall Sensing Range

In the LOW range

One yellow LED lights up

In the MEDIUM range

Two yellow LEDs light up

In the HIGH range

Three yellow LEDs light up

The limit value cannot be set to less than or equal to the currently selected reset hysteresis!

DELAY TIME "TD"

Adjustable from 0.1 to 9.9 seconds (default value: 0.1 second).

RESET HYSTERESIS "HY"

Setting range: 1 to 25% of the selected limit value.
The output relays are not switched back until the measured value decreased below the selected percentage of hysteresis

Reset hysteresis cannot be set to greater than or equal to the currently selected limit value!

The default setting for reset hysteresis is 1%.

TEMPERATURE MONITORING "TP" (TEMPERATURE PROTECTION)

Setting range: on/off
The connected probe is designed for a specific temperature range. If the permitted range is exceeded, this is indicated as an error. If this alarm

function is not desired, temperature monitoring can be deactivated (Off).

TUR

19-08-2021

M-410.03-EN-AA

410-03

/2

Turbidity Limit Value

Evaluation Unit

TURBISWITCH GS5

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Содержание TURBISWITCH GS5

Страница 1: ...OMAT series TT GS TR GS and CP1 at the evaluation unit TURBISWITCH GS5 is not possible TECHNICAL DATA Supply power 100 240V AC 50 60Hz TURBISWITCH GS5 G or 10 30V DC and 12 24V AC TURBISWITCH GS5 D Po...

Страница 2: ...shed by simply turning the rotary pushbutton selector The selected hysteresis value is automatically skipped in this case 100 in the LOW range is followed by 0 in the MEDIUM range 100 in the MEDIUM ra...

Страница 3: ...iefly displayed The display is then limit value display INSTALLING A MEASURING CELL IN A PIPING SYSTEM IN LINE Installation of in line measuring cell The Measuring cells TURBISWITCH GA1 GA5 GA11 can b...

Страница 4: ...TURBISWITCH CP2 and CP5 transducers are designed as hanging probes or for use with an extension tube It can be installed in slurry tanks basins or open wells The probe should be installed so that it c...

Страница 5: ...eiver TT HDR and TR HDR or TT HDR5 and TR HDR5 Emitter Receiver IR Light TT HDR Emitter TURBISWITCH GS5 TR HDR Receiver 10 to 30 V DC 100 to 240 V AC L N Fuse 16 14 8 2 6 1 3 4 OC V RX V Limit value L...

Страница 6: ...13 15 Switching logic with increased reliability The contacts open when one of the following events occur Contact 9 11 opens when the limit value is exceeded Contact 9 12 opens when measurements are b...

Страница 7: ...NG THE THRESHOLD LIMIT VALUE AND COMMISSIONING Adjustment takes place after installation and electrical connection If problems arise see Troubleshooting All steps must be carried out in chronological...

Страница 8: ...is so high that no light is reaching the receiver HDR5 emitter blinks red Defective emitter Replace the emitter TR appears on the display The receiver has malfunctioned or it is not connected Check t...

Страница 9: ...mescale remover The side of the glass disc marked with a dot must be installed that it faces the medium Check the seals O rings for damage and replace them if necessary before start up The frequency o...

Отзывы: