BALLANTINE LABORATORIES, Inc.
MODEL 3207 MICRO-OHM METER
30
SECTION 6
SPECIFICATIONS
6.1 Electrical:
Range
(Ohm)
Resolution
(Ohm)
Accuracy
+ (% rdg + dgt)
FS rdg.
(Ohm)
DC Current
applied to
DUT
Power
dissipation
in the
sample at
mid-scale
2m 0.1µ +(0.5+5) 1.9999m 1A
1mW
20m 1µ +(0.05+5) 19.999m 1A
10mW
200m 10µ +(0.05+5) 199.99m 100
mA 1mW
2000m 100µ +(0.05+5) 1999.9m 10
mA 0.1mW
20 1m +(0.05+5) 19.999 10mA 1mW
200 10m +(0.05+5) 199.99 1mA 0.1mW
2000 100m +(0.05+5) 1999.9 0.1mA 0.01mW
Note: Accuracies are valid from 1/10
th
of the range to FS of range and valid at.
Temperature: 25ºC
±
3ºC at 55%
±
10 % RH.
Tempco. : 150 ppm/ºC.
Warm up time for stated accuracy – 5 minutes
Over range – blinking Zero display.
6.2 Environmental :
Temperature:
Operating : 0ºC to 40ºC
Storage : 0ºC to 50ºC
Humidity :
<
90% RH Non Condensing.
6.3
General :
Power
: 115 VAC ± 10%, 60Hz, single phase (220VAC on order)
Dimensions
: 340 (D) x 258(W) x 145(H) mm (approx.)
Net Weight
: 3.2 Kg.(approx.) (4.0 Kg. for Option –05)
Shipment Weight
: 5.2 Kg. (approx.)