COMMAND TOKEN
ACTION PERFORMED
CHARACTER
ASCII VALUE in HEX
‘0’ 30
Selects 2m
Ω
Range
‘1’ 31
Selects 20m
Ω
Range
‘2’ 32
Selects 200m
Ω
Range
‘3’ 33
Selects 2000m
Ω
Range
‘4’ 34
Selects 20
Ω
Range
‘5’ 35
Selects 200
Ω
Range
‘6’ 36
Selects 2000
Ω
Range
‘i’ OR ‘I’
69 OR 49
Increments the range
‘d’ OR ‘D’
64 OR 44
Decrements the range
‘r’ OR ‘R’
72 OR 52
Reads the range selected returning the
ASCII value from ‘0’ to ‘6’ of the range.
‘t’ OR ‘T’
74 OR 54
Reads the 5 digit (Max) count *
‘o’ OR ‘O’
6F OR 4F
Operate Mode
‘s’ OR ‘S’
73 OR 53
Stand-By mode
*Interpretation of count
For example: If the count is 668,then read the range. If the range is
‘2’ (range=200m
Ω
), then read the count as 6.68 m
Ω
.
No.
ASCII
character
Range selected Maximum display
1 ‘0’
2m
Ω
1.9999
2 ‘1’
20m
Ω
19.999
3 ‘2’ 200m
Ω
199.99
4 ‘3’ 2000m
Ω
1999.9
5 ‘4’
20
Ω
19.999
6 ‘5’
200
Ω
199.99
7 ‘6’ 2000
Ω
1999.9
Содержание 3207
Страница 2: ......
Страница 4: ......
Страница 6: ...BALLANTINE LABORATORIES Inc MODEL 3207 MICRO OHM METER 6 ...
Страница 10: ...BALLANTINE LABORATORIES Inc MODEL 3207 MICRO OHM METER 10 ...
Страница 12: ...BALLANTINE LABORATORIES Inc MODEL 3207 MICRO OHM METER 12 ...
Страница 20: ...BALLANTINE LABORATORIES Inc MODEL 3207 MICRO OHM METER 20 FIG 3 1 ...
Страница 35: ...BALLANTINE LABORATORIES Inc MODEL 3207 MICRO OHM METER 35 ATTACHMENT CALIBRATION FORM ...