Overview and Safety Instructions
1.
Please read this manual carefully to familiarise yourself with the operation and maintenance of your
Invivo
2
Plus workstation.
Note
: The workstation should be located in a well-ventilated area.
For your safety, safety of others using the workstation and those around you;
•
The covers on both ends and the top of the workstation must not be removed by anyone other
than a qualified service engineer. There are no end user serviceable parts within these covers.
•
The AC Mains outlet that the Invivo
2
Plus workstation is connected to, MUST not be
obstructed by the Invivo
2
Plus workstation or any other equipment, and MUST be accessible in
case of emergency. In case of emergency, disconnect the Invivo
2
Plus workstation from the AC
Mains Outlet.
•
In case of damage to the Invivo
2
Plus workstation, disconnect the Invivo
2
Plus workstation
from the AC Mains Outlet and contact your local distributor for advice.
•
The workstation must be connected to a protective earth.
•
Only the power cord supplied with the workstation should be used to connect the workstation
to the mains supply.
•
Only Ruskinn Technology Limited replacement parts should be used.
•
Gas regulators must be used for each gas supply. A 2 stage regulator should be used for a
bottled gas supply. The maximum supply pressure permissible is 4 bar gauge.
•
The maximum permissible concentration of Hydrogen in the anaerobic mixed gas is 5.5%
Hydrogen.
•
The mains supply voltage fluctuations must not /- 10% of the nominal mains voltage.
•
Only the gasses specified in this manual may be used.
•
The maximum power ratings of the internal sockets must not be exceeded.
•
The exhaust valve outlet must not be covered or blocked.
•
The cooling fan covers and vents must not be covered or blocked.
•
The workstation must be disconnected from the mains supply before removing the front
screen. The workstation must not be reconnected to the mains supply until the front screen
has been reinstalled.
•
The workstation should not be lifted by the glove ports, the interlock or the Single Plate Entry
System.
•
If an Uninterruptable Power Supply (UPS) system is used, both the Invivo
2
Plus workstation
and the ICO
2
N
2
IC Advanced Gas mixing system must be connected to the Uninterruptable
Power Supply (UPS) system.
•
The weight limit for the interlock floor tray is 4kg.
•
The weight limit for the rear shelf is 12.5kg, evenly distributed.
•
The spot light should not be used continuously for a period of more than 10 minute. The spot
light should be allowed to cool for a period of 10 minutes after each use.
•
The use of Radioactive materials in the Invivo
2
400 is strictly prohibited.
FAILURE TO ADHERE TO THESE SAFETY INSTRUCTIONS COULD CAUSE SERIOUS INJURY AND WILL
INVALIDATE THE WORKSTATION WARRANTY. RUSKINN TECHNOLOGY LIMITED ACCEPTS NO
RESPONSIBILITY FOR ANY ACCIDENT, INJURY OR LOSS CAUSED BY UNSAFE OPERATION OF THE
WORKSTATION