Baker Invivo2 300 UM-010 Скачать руководство пользователя страница 45

Invivo

300

 

 

 

 

 

UM-010 R4.0 

Page 45 of 48 

Invivo

2

 300 User Manual 

 

Gas mixer and gas consumption problems 

The Invivo

2

 300 workstation is fitted with a gas demand indicator (item 7 in Figure 9). When the gas 

demand indicator is illuminated, the pressure is low in the workstation chamber. A shot of gas will be 
injected into the workstation to raise the pressure in the workstation chamber. In normal operation, 
the  gas  demand  indicator  should  not  illuminate  more  than  once  every  20  minutes  when  the 
workstation is at rest. At rest is defined as the gasses are within Deadband and glove ports and SPES 
are closed. Figure 37 gives a list of common problems related to the gas mixer and gas consumption; 

Figure 37: Gas Mixer workstation related and gas consumption common problems 

Problem 

Potential Causes 

Solution 

The gas demand indicator 
illuminates  more  than 
once  every  20  minutes 
whilst  at  rest/  high  gas 
consumption 
 

 

The Ezee Sleeves  are not 
correctly installed 
 
The glove port cover is not 
correctly closed  
 
The  cuff  is  not  tight 
around the users wrist 
 
 
 
The  humidifier  tank  is 
empty 
 
A gas cylinder is empty 
 
 
 
 
The  interlock  inner  and 
outer  doors  are  open  at 
the same time 
 
The  SPES  has  been  left 
open 
 
The 

workstation 

has 

developed a leak 

Check the Ezee Sleeves are correctly installed. See 
section 0 for more details 
 
Check the glove port cover is in place and closed 
correctly. Reclose the glove port cover if required 
 
Ensure that the correct size Ezee Sleeve is used. 
Ensure  that  the  cuff  is  pulled  tight  around  the 
wrist.  Roll  up  sleeves  of  clothing  and  remove 
jewellery 
 
Refill  the  humidifier  tank.  See  section  7.2.1,  for 
more details 
 
Check the ICO

2

N

2

IC gas mixer screen to see which 

gas cylinder is empty. An empty cylinder  will be 
indicated  by  a  low  pressure  status.  Replace  the 
gas cylinder with a full cylinder 
 
Close  either  the  interlock  inner  door  or  the 
interlock outer door 
 
 
Close the SPES. See section 6.3 for more details 
 
 
Contact your local distributor 
 

In  anoxic  level mode,  the 
oxygen 

concentration 

does not reach 0.0% 
 

There is no catalyst sachet 
installed 
 
The catalyst sachets need 
replacing 

Install  a  catalyst  sachet.  See  section  7.2.4  for 
more details. 
 
Replace the catalyst sachets. See section 7.2.4 for 
more details. 

The  gas  mixer  alarm  is 
sounding 
 

A gas cylinder is empty 

Check the ICO

2

N

2

IC gas mixer screen to see which 

gas cylinder is empty. An empty cylinder  will be 
indicated  by  a  low  pressure  status.  Replace  the 
gas cylinder with a full cylinder 

For all other ICO

2

N

2

IC Gas Mixer related problems, please refer to UM-011 ICO

2

N

2

IC user manual. 

 

Содержание Invivo2 300 UM-010

Страница 1: ...Invivo2 300 UM 010 Invivo2 300 User Manual Affix Serial Number Sticker Here...

Страница 2: ...transfer 18 plates in 35 seconds Utilised with the ICO2N2IC Advanced Gas Mixing System to offer a microaerophilic environment in which to grow facultative anaerobes Optimum cell environment o Accurat...

Страница 3: ...rements 10 Voltage and frequency requirements 10 Power consumption 10 Powering the workstation 11 Workstation overview 12 Workstation layout 12 Component layout 12 Front view 12 Rear view 13 Left side...

Страница 4: ...ccess to the main chamber 24 Ezee Sleeve overview 24 Entering the workstation chamber using the Ezee Sleeves 25 Exiting the workstation chamber using the Ezee Sleeves 26 Arm gas displacement volume an...

Страница 5: ...38 Replacing the mains plug fuse UK users only 39 Replacing the mains fuses 40 Service requirements 41 Spare parts and accessories 41 Overview 41 Spare parts and cleaning agents 42 Accessories 42 Wor...

Страница 6: ...ible is 4 bar gauge The maximum permissible concentration of Hydrogen in the anaerobic mixed gas is 5 5 Hydrogen The mains supply voltage fluctuations must not exceed 10 of the nominal mains voltage O...

Страница 7: ...on should be located in a well ventilated area Figure 1 Invivo2 300 symbols Symbol Meaning Refer to user manual Alternating current O Off I On Primary earth connection This product complies with the e...

Страница 8: ...ardous components and must not be disposed of at a household waste site Instead it should be taken to the appropriate collection point for the recycling of electrical and electronic equipment Date of...

Страница 9: ...External height 650 mm External depth 660 mm Workstation chamber internal width 500 mm Workstation chamber internal height 420 mm Workstation chamber internal depth 460 mm Interlock internal width 15...

Страница 10: ...ernal components of the workstation and will invalidate the warranty The gasses are connected to the ICO2N2IC Advanced Gas Mixer See UM 011 ICO2N2IC user manual for more details Electrical supply requ...

Страница 11: ...tions inside the room the workstation is located in Powering the workstation To switch the workstation on press the power switch item 2 Figure 14 To switch the workstation off press the power switch i...

Страница 12: ...inner and an outer door The Invivo2 300 workstation is controlled via the control panel The ICO2N2IC Gas Mixer controls the gaseous composition of the internal environment See UM 011 for more informa...

Страница 13: ...Invivo2 300 UM 010 R4 0 Page 13 of 48 Invivo2 300 User Manual Rear view Figure 5 shows the rear view of the Invivo2 300 workstation Figure 5 Invivo2 300 workstation rear view...

Страница 14: ...workstation left side view 1 Pressure relief tank cover pressure relief tank bung 2 Pressure relief tank 3 Pressure relief tank overflow pipe 4 Cable gland optional extra 5 Exhaust valve outlet 6 Vacu...

Страница 15: ...vo2 300 workstation with the ultrasonic humidifier Figure 7 Invivo2 300 workstation left side view with ultrasonic humidifier 1 Exhaust valve outlet 2 Cable gland optional extra 3 Humidifier screw cap...

Страница 16: ...8 shows the right side view of the Invivo2 300 workstation Figure 8 Invivo2 300 workstation right side view 1 Mixed gas inlet 2 Sample gas return 3 Sample gas out 4 Serial number label 5 Portable Appl...

Страница 17: ...switched on 2 Chamber light switch Press to switch the chamber light on off 3 Gas Alarm switch Press to switch the gas alarm on off 4 Humidistat Controls the level of humidity inside the workstation c...

Страница 18: ...erature 0 1 C increments or humidity 5 Increase temperature 0 1 C increments or humidity For temperature control the white value PV Process Value is the actual temperature The green value SV Set Value...

Страница 19: ...nates to indicate when the interlock inner door is available 2 Inner door button Press to unlock the interlock inner door when available Foot switch control panel Figure 13 shows the foot switch contr...

Страница 20: ...Manual Power entry panel Figure 14 shows the power entry panel and label Figure 14 Power entry panel 1 Mains fuse holder 2 On off switch 3 Mains lead connection 4 Nitrogen inlet 5 Primary earth stud...

Страница 21: ...of humidity control on Invivo2 workstations Standard Ultrasonic Humidistat is the standard method of humidity control on the Invivo2 300 workstation Ultrasonic is an optional extra Note Ultrasonic hum...

Страница 22: ...pening either the outer or inner door Note Both doors should not be opened at the same time Opening the interlock outer door To open the interlock outer door pull the interlock door towards you using...

Страница 23: ...the foot switch control panel item 2 Figure 13 The interlock purge cycle takes 30 seconds Access the workstation chamber via the Ezee Sleeves See section 6 2 2 for details When the interlock purge cy...

Страница 24: ...into the workstation chamber via the SPES and unloading the interlock Right hand glove port only for unloading the interlock For each method of entry the glove port selector switch Figure 11 must be...

Страница 25: ...the Ezee Sleeves Repeat the procedure for the other hand if both hands are to enter the workstation chamber Press and hold the vacuum pedal with your foot item 1 in Figure 13 until the internal gas v...

Страница 26: ...e port bar inside the glove port so that it is vertical Tighten the glove port cover knob Do not over tighten the glove port cover knob Supporting the Ezee Sleeve cuff with your opposite hand pull you...

Страница 27: ...known as the mailbox SPES overview The SPES consists of an external hinged flap and an internal hinged flap The external flap is held in place by a thumb screw when not in use The internal flap is se...

Страница 28: ...back to its closed position Whilst holding the SPES external flap swing the thumb screw right to slot into the SPES external flap Tighten the thumb screw Note Do not over tighten the thumb screw Note...

Страница 29: ...tem 3 in Figure 13 To switch the spotlight off remove your foot from the spot pedal on the foot switch control panel item 3 in Figure 13 Note The spot light should not be used continuously for a perio...

Страница 30: ...e gland is suitable for cables of diameter 3 5mm to 7mm The cable gland port is located on the left hand side as shown in Figure 6 item 3 Figure 23 shows the cable gland Figure 23 Cable gland To use t...

Страница 31: ...ess media from Petri dishes The vacuum port is located on the left hand side of the workstation To use the vacuum port Connect the vacuum source to the external part of the vacuum port by pushing a tu...

Страница 32: ...d demonstrations can be found at Ruskinn Technology Limited s YouTube channel http www youtube com ruskinntechnology Video demonstrations are available for Entering and exiting the workstation chamber...

Страница 33: ...mer Tristel Fuse Sachets and Duo Foamer are available from Ruskinn Technology Limited see section 7 4 2 for details Ruskinn Technology Limited anti static cleaner Distilled water No other cleaning age...

Страница 34: ...e of the SPES o The glove port covers For all workstation components wipe with a paper towel soaked in an appropriate cleaning agent Take care not to get cleaning agents on the fans below the inner re...

Страница 35: ...een see UM 011 for more information Either open the SPES see section 6 3 2 or open one of the glove port covers see section 6 2 2 to equalise the pressure between the workstation chamber and the exter...

Страница 36: ...ows the O ring removal Figure 28 O ring removal Remove the Ezee Sleeve To install an Ezee Sleeve Place the new Ezee Sleeve onto the glove port ensuring that the O ring grooves are completely covered a...

Страница 37: ...e Ezee Sleeves see section 0 for more information Undo the glove port covers and place on the rear shelf Lift the floor tray at the front to expose the detox sachets Figure 29 shows the location of th...

Страница 38: ...orage facility Return the ICO2N2IC to the main menu Switch off Invivo2 300 and disconnect from the mains power supply Remove the Ezee Sleeves see section 0 for more information Undo the glove port cov...

Страница 39: ...from the mains socket Using a small flat bladed screw driver remove the fuse cover from the mains plug Figure 30 shows the fuse removal Figure 30 Mains plug fuse removal Replace the fuse with a BS 136...

Страница 40: ...removal of the mains fuse holder Figure 31 mains fuse holder removal Replace the mains fuses The fuse ratings are o 240V 50Hz F3 15A H250V o 220V 60Hz F3 15A H250V o 110V 60Hz F5A H250V o 100V 50 60Hz...

Страница 41: ...ineer Biennial service Biennially 2 yearly Qualified service engineer To arrange an annual or biennial service contact your local distributor Note that the biennial service includes an annual service...

Страница 42: ...tion Anaerobic Indicator Strips Anoxic operation Accessories A range of accessories are available to enhance the functionality of your Invivo2 300 workstation Figure 34 is the list of accessories for...

Страница 43: ...he workstation is not plugged in The mains socket is not switched on The mains plug fuse has blown The mains fuse has blown Plug the workstation into the mains Switch the mains power on UK users Repla...

Страница 44: ...rlock inner door and remove them Wait for the interlock purge cycle to finish Spillages can cause the interlock inner door to stick Clean the interlock inner door See section 7 1 2 7 1 3 for more deta...

Страница 45: ...s been left open The workstation has developed a leak Check the Ezee Sleeves are correctly installed See section 0 for more details Check the glove port cover is in place and closed correctly Reclose...

Страница 46: ...ARRANTY OF FITNESS FOR A PARTICULAR PURPOSE THE IMPLIED WARRANTY OF MERCHANTABILITY OR ANY IMPLIED WARRANTY ARISING OUT OF A COURSE OF DEALING OR OF PERFORMANCE CUSTOM OR USAGE OF TRADE Ruskinn Techno...

Страница 47: ...not be disposed of at a household waste site Instead it should be taken to the appropriate collection point for the recycling of electrical and electronic equipment Alternatively please contact your l...

Страница 48: ...les sales ruskinn com Technical support techsupport ruskinn com General enquiries ruskinnoffice ruskinn com Website www ruskinn com YouTube channel http www youtube com ruskinntechnology Ruskinn Techn...

Отзывы: