Lower the floor tray, ensuring that the floor tray leg has located in the floor tray locator (item
3 in Figure 29).
Close the glove ports using the glove port covers.
Reinstall the Ezee Sleeves (see section 0 for more details).
Reconnect the Invivo
2
300 to the mains power supply and switch the Bugbox Plus on.
Installing/ replacing a catalyst sachet
Catalyst sachets are available as an accessory for the Invivo
2
300 workstation. See section 7.4.2 for
ordering details. The catalyst sachets are only required for anaerobic operation of the workstation.
Anaerobic mode is an optional extra for the ICO
2
N
2
IC gas mixer.
To install a catalyst sachet;
Remove any samples to an alternative storage facility.
Return the ICO
2
N
2
IC to the main menu.
Switch off Invivo
2
300 and disconnect from the mains power supply.
Remove the Ezee Sleeves (see section 0 for more information).
Undo the glove port covers and place on the rear shelf.
Lift the floor tray at the front to expose the detox sachets.
Slide a sachet forward to remove it from its holder.
Remove the sachet via the glove port.
Lower the floor tray, ensuring that the floor tray leg has located in the floor tray locator (item
3 in Figure 29).
Remove the catalyst sachet from its packaging.
Whilst lifting the floor tray at the front, slide the catalyst sachet into the sachet holder.
Lower the floor tray, ensuring that the floor tray leg has located in the floor tray locator (item
3 in Figure 29).
Close the glove ports using the glove port covers.
Reinstall the Ezee Sleeves (see section 0 for more details).
Reconnect the Invivo
2
300 to the mains power supply and switch on the Invivo
2
300.