Invivo
2
300 User Manual
Cleaning procedure – deep clean
To deep clean the workstation;
Preparing the workstation
Return the ICO
2
N
2
IC to the main menu (see UM-011 for more information).
Remove all cells/ samples to an alternative storage facility.
Switch the workstation off at the mains and remove the plug from the mains.
Remove the Ezee Sleeves
(see section 0, for more details).
Remove any other equipment installed in the workstation chamber.
Cleaning the workstation chamber
The items inside the workstation chamber that require cleaning are;
o
The floor tray.
o
The ceiling panel.
o
The shelf panel.
o
The left hand wall.
o
The right hand wall.
o
The interlock inner door.
o
The inside of the SPES.
o
The glove port covers.
For all workstation components, wipe with a paper towel soaked in an appropriate cleaning
agent. Take care not to get cleaning agents on the fans (below the inner rear wall), plug socket
and the interlock inner door control panel.
Wipe dry using paper towels.
Cleaning the interlock chamber;
The items inside the interlock chamber that require cleaning are;
o
The floor.
o
The ceiling panel.
o
The rear wall.
o
The right hand wall.
o
The interlock inner door.
For all workstation components, wipe with a paper towel soaked in an appropriate cleaning
agent. Take care not to get cleaning agents on the interlock heating mat (below the interlock
floor). Note the interlock floor can be removed for cleaning.
Wipe dry using paper towels.
Reinstalling workstation components
Reinstall any equipment removed from the workstation chamber.
Reinstall the Ezee Sleeves
(see section 0, for more details).