background image

Options and Accessories

For convenience, most options, accessories and modifications are factory installed and should be specified when 
ordering.  Commonly requested options are listed below. For detailed information on accessories and modifications, 
contact The Baker Company.

• 

Additional service fixtures (air, gas, vacuum, cold or DI water, steam)

• 

Cold water gooseneck faucet

• 

Distilled/DI water gooseneck

• 

Vacuum breakers

• 

Aspirator

• 

Epoxy resin sink

• 

Base cabinet (see below)

• 

Stainless steel peg board

• 

Tubular base stand

• 

Vertical only sash

• 

Cup sink

• 

Stainless steel work surface

PAGE 

6

Technical Specifications

MODEL NUMBER  

FH4 

FH5 

FH6 

Dimensions

  Nominal Size 

4' 

5' 

6' 

  Fume Hood (W x D x H) 

48” x 33 ¾” x 59” 

60” x 33 ¾ ” x 59” 

72” x 33 ¾” x 59" 

   

[1,219 x 857 x 1,499 mm] 

 [1,524 x 857 x 1,499 mm] 

[1,829 x 857 x 1,499 mm]

  Base Cabinet (W x D x H) 

48” x 29” x 35”  

60” x 29”x 35 ¾”  

72” x 29”x 35 ¾” 

   

[1,219 x 737 x 908 mm] 

[1,524 x 737 x 908 mm] 

[1,829 x 737 x 908 mm]

  Interior Dimensions (W x D x H) 

38 5⁄8" x 18 ½"x 24 7⁄16”  

50 5⁄8" x 18 ½" x 24 7⁄16”  

62 5⁄8" x 18 ½" x 24 7⁄16”  

   

[981 x 470 x 621 mm] 

 [1,286 x 470 x 621 mm]  

[1,591 x 470 x 621 mm]

  Useable Work Surface (W x D) 

38 5⁄8"

 

x 18 ½” 

50 5⁄8" x 18 ½” 

62 5⁄8" x 18 ½” 

   

[981 x 470 mm] 

[1,286 x 470 mm] 

[1,591 x 470 mm]

Exhaust and Static Pressure Requirements - Combination Horizontal/Vertical Sash

 

  Exhaust Airflow Rate (vertical sash at max width) 

865 CFM [408 L/Sec] 

1204 CFM [568 L/Sec] 

1286 CFM [607 L/Sec]

  Exhaust Airflow Rate (horizontal sash at max width) 

374 CFM [164 L/Sec]  

405 CFM [191 L/Sec]  

623 CFM [293 L/Sec] 

Electrical

  Electrical Service Requirements 

115V AC, 15A, 60 Hz 

115V AC, 15A, 60 Hz 

115V AC, 15A, 60 Hz

  Operating Amperage 

1.1 A 

1.1 A 

1.2 A

  Power Consumption 

126.5 W 

126.5 W 

138 W

  Heat Generation 

431.6 Btu/Hr  

431.6 Btu/Hr  

470.9 Btu/Hr

  Outlets 

Two GFCI protected, 115 V AC, Duplex outlets. The outlets on this circuit are protected by a self-resetting  

   

circuit breaker prewired to a junction box. This breaker allows a total of 10 A on all outlets.

Содержание ChemGARD FH4

Страница 1: ...Underwriter Laboratories Listed C h e m G A R D ...

Страница 2: ...ces roll effect fluctuations in the airflow that may cause harm to the worker Offers a streamlined unobstructed work area Minimizes lingering concentrations above and behind the viewscreen Reduces contaminant concentrations near the edge of the sash window reducing potential exposure hazard to personnel ChemGARD FumeHoods ...

Страница 3: ...alarm monitor configurations 4 Angled entrance airfoil and radius work surface for undisturbed airflow into work area 5 Color coded service fixtures with convenient remote control valves positioned outside the air stream 6 Counterbalanced vertical sash for smooth single handed operation 7 Recessed work surface confines spills 8 All stainless steel construction for durability and easy cleaning 9 Op...

Страница 4: ...el liner provides long term durability easy cleaning and resistance to corrosion can be epoxy coated as an option A molded dished epoxy resin work surface confines spills Recessed work surface encourages proper placement of potentially hazardous chemicals ChemGARD RI Radioisotope Models Formed radii precision engineered 16 gauge Type 304 stainless steel coved corners water tight welded and polishe...

Страница 5: ...ed controls include flush membrane switches that are easy to operate and easy to clean Independent light and duplex electrical outlet switches are included on the main control panel Remote alarm contacts allow integration with existing laboratory monitoring systems Horizontal VerticalSashOffersAdvantages The ChemGARD incorporates a combination horizontal vertical sash designed for energy savings o...

Страница 6: ...atures include Acid Storage Base Cabinet The acid storage base cabinet has an acid resistant polypropylene drip tray shelf liner and optional vent assembly with flexible polypropylene hose designed to T into the fume hood exhaust plenum Reinforced shelf adjustable on 1 centers Dual panel doors 1 thick secured with 3 wide continuous hinge 3 point latch 18 gauge cold rolled steel bottom pan recessed...

Страница 7: ...ll andtwofrontremovablerearpanelsprovideaccesstothe10 deep chase at the rear to allow for installation of the optional vent package Access panels are located at each end of the chase An optional vent package offers from 10 to 15 air changes per hour depending on cabinet size the vent assembly with flexible hose is designed to T into the fume hood A reinforcedType 304 stainless steel shelf is adjus...

Страница 8: ...x 908 mm 1 524 x 737 x 908 mm 1 829 x 737 x 908 mm Interior Dimensions W x D x H 38 5 8 x 18 x 24 7 16 50 5 8 x 18 x 24 7 16 62 5 8 x 18 x 24 7 16 981 x 470 x 621 mm 1 286 x 470 x 621 mm 1 591 x 470 x 621 mm Useable Work Surface W x D 38 5 8 x 18 50 5 8 x 18 62 5 8 x 18 981 x 470 mm 1 286 x 470 mm 1 591 x 470 mm ExhaustandStaticPressureRequirements CombinationHorizontal VerticalSash Exhaust Airflo...

Страница 9: ......

Страница 10: ...the invoice date The exclusive remedy for any breach or violation of this warranty is as follows The Baker Company Inc will F O B Sanford Maine furnish without charge repairs to or replacement of the parts or equipment that proved defective in material or workmanship No claim may be made for any incidental or consequential damages THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES EXPRESSE...

Отзывы: