background image

Lighting and Electrical System

• 

A sealed, fluorescent light fixture is located above gasketed safety-plate glass; fluorescent lamps are replaced from 
the front.

• 

Solid-state lamp ballasts reduce heat output and improve efficiency.

• 

Two GFCI duplex outlets are protected by a single 10-amp self resetting circuit breaker to protect personnel and 
minimize tripping of breakers.

Utilities

• 

Color-coded service fixtures for air, gas, vacuum, cold or DI water and steam, include convenient remote control 
valves.

• 

The  cabinet  includes  one  fixture  and  one  petcock  standard,  plus  seven  plugged  penetrations  for  additional  
optional fixtures.

• 

Pre-studded exhaust outlet, 10" diameter, accepts flanged exhaust connection.

• 

Airwatch II airflow monitor.

General Purpose Base Cabinet

The general purpose base cabinet has removable front and side panels, plus a 10" deep chase at the rear to allow room 
for services. Other features include:

Acid Storage Base Cabinet

The acid storage base cabinet has an acid-resistant polypropylene drip tray, shelf liner and (optional) vent assembly 
with flexible polypropylene hose designed to “T” into the fume hood exhaust plenum.

• 

Reinforced shelf, adjustable on 1" centers.

• 

Dual panel doors, 1 ½" thick, secured with 3" wide continuous hinge, 3-point latch.

• 

18-gauge, cold-rolled steel bottom pan, recessed 1½", liquid tight. 2½" riser with 3⁄8" leveling feet.

• 

16-gauge, cold-rolled steel side panels.

• 

Epoxy powder coated throughout.

• 

2" diameter plugged penetrations in back wall, and removable side and rear panels with a 10" deep chase at the 
rear to allow for installation of the optional vent package.

• 

The vent package offers from 10 to 15 air changes per hour, depending on cabinet size.

• 

Reinforced steel shelf, adjustable on 1" centers.

• 

Dual panel doors, 1½" thick, secured with 3" wide continuous hinge, 3-point latch.

• 

18-gauge, cold-rolled steel bottom pan, recessed 1½", liquid tight. 2½" riser with 3⁄8" leveling feet. 16-gauge, cold-
rolled steel side panels.

• 

Epoxy powder coated throughout.

Содержание ChemGARD FH4

Страница 1: ...Underwriter Laboratories Listed C h e m G A R D ...

Страница 2: ...ces roll effect fluctuations in the airflow that may cause harm to the worker Offers a streamlined unobstructed work area Minimizes lingering concentrations above and behind the viewscreen Reduces contaminant concentrations near the edge of the sash window reducing potential exposure hazard to personnel ChemGARD FumeHoods ...

Страница 3: ...alarm monitor configurations 4 Angled entrance airfoil and radius work surface for undisturbed airflow into work area 5 Color coded service fixtures with convenient remote control valves positioned outside the air stream 6 Counterbalanced vertical sash for smooth single handed operation 7 Recessed work surface confines spills 8 All stainless steel construction for durability and easy cleaning 9 Op...

Страница 4: ...el liner provides long term durability easy cleaning and resistance to corrosion can be epoxy coated as an option A molded dished epoxy resin work surface confines spills Recessed work surface encourages proper placement of potentially hazardous chemicals ChemGARD RI Radioisotope Models Formed radii precision engineered 16 gauge Type 304 stainless steel coved corners water tight welded and polishe...

Страница 5: ...ed controls include flush membrane switches that are easy to operate and easy to clean Independent light and duplex electrical outlet switches are included on the main control panel Remote alarm contacts allow integration with existing laboratory monitoring systems Horizontal VerticalSashOffersAdvantages The ChemGARD incorporates a combination horizontal vertical sash designed for energy savings o...

Страница 6: ...atures include Acid Storage Base Cabinet The acid storage base cabinet has an acid resistant polypropylene drip tray shelf liner and optional vent assembly with flexible polypropylene hose designed to T into the fume hood exhaust plenum Reinforced shelf adjustable on 1 centers Dual panel doors 1 thick secured with 3 wide continuous hinge 3 point latch 18 gauge cold rolled steel bottom pan recessed...

Страница 7: ...ll andtwofrontremovablerearpanelsprovideaccesstothe10 deep chase at the rear to allow for installation of the optional vent package Access panels are located at each end of the chase An optional vent package offers from 10 to 15 air changes per hour depending on cabinet size the vent assembly with flexible hose is designed to T into the fume hood A reinforcedType 304 stainless steel shelf is adjus...

Страница 8: ...x 908 mm 1 524 x 737 x 908 mm 1 829 x 737 x 908 mm Interior Dimensions W x D x H 38 5 8 x 18 x 24 7 16 50 5 8 x 18 x 24 7 16 62 5 8 x 18 x 24 7 16 981 x 470 x 621 mm 1 286 x 470 x 621 mm 1 591 x 470 x 621 mm Useable Work Surface W x D 38 5 8 x 18 50 5 8 x 18 62 5 8 x 18 981 x 470 mm 1 286 x 470 mm 1 591 x 470 mm ExhaustandStaticPressureRequirements CombinationHorizontal VerticalSash Exhaust Airflo...

Страница 9: ......

Страница 10: ...the invoice date The exclusive remedy for any breach or violation of this warranty is as follows The Baker Company Inc will F O B Sanford Maine furnish without charge repairs to or replacement of the parts or equipment that proved defective in material or workmanship No claim may be made for any incidental or consequential damages THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES EXPRESSE...

Отзывы: