4
SAFETY INSTRUCTIONS
Baker and/or Ruskinn do not take any responsibility for damages caused by using the equipment for other
purposes than described in this user manual.
This is a Class A product (for industrial environments). In a residential environment, it may cause radio
interference. If radio interference occurs, the user may be required to take appropriate measures.
This device contains high pressure gases.
If there is gap between gas bottle and regulator do not use or fill this bottle. Place this bottle on ground
on an isolated safe place to let the gas bottle to fully degas.
This device contains electronics.
This device contains heating elements.
Do not change heating device type settings from heat plate controller.
Do not replace any parts of the device or its components.
This device is intended to use in research only.
Device should be only used by technically qualified person.
Device must be used according to manual instructions, if usage deviates from the instructions the
performance can be harmed.
If any damages appear, the device should not be used.
If unusual performance occurs, such as deformation, discoloration, abnormal sound or smell, or abnormal
heating stop using the device and contact the provider.
If any liquid leaks from interior battery do not touch the fluid.
Always transport the device with care and do not drop or apply any shock or vibrations to the device.
Always place device on robust and stable surface.
Some parts of the device can reach high temperatures, thus do not touch these parts.
Avoid rapid temperature changes, moisture, moist condensation, air drafts, direct sunlight, temperature
way above or below room temperature, excessive induction noise, static electricity and magnetic fields, in
the device surroundings and direct contact.
Do not use this device in same surroundings with flammable or explosive gases.
Do not disassemble any parts from device.
Do not use harsh solvents to clean the instrument, this may cause disruption of device.
Do not induce the device for high voltages and use only electric cords equipped with the device.
Do not induce physical wear to cables, tubes or any parts.
Avoid dust particles near gas bottle connections.
Use only the power supply provided with device.
Always connect power supply cable first to the device before connecting it to a power socket.
Do not connect power supply cable or turn on the device if any cables are damaged.
Connect the device only to grounded main power distribution network.
When handling gas bottles be extra careful.
Always check your gas bottles before using.
Do not use or fill gas bottles that are damaged.
Do not fill bottles with gas mixtures containing oxygen more than 21% at any circumstances.
Under NO CIRCUMSTANCES should the system be filled with pure oxygen. Oxygen will ignite causing injury
or death.
Do not let any dirt, water or oils to become in contact with gas bottle regulators.