35
Settings: Alarms
Alarm can be set to inform if temperature
deviates from set point for defined period of
time. Alarm can be beside visual also audible.
Click alarms
icon to set alarm:
1.
Define the time that heat plate temperature
may remain outside allowed tolerance
before triggering alarm.
2.
Define
the
tolerance
value
which
temperature cannot go over or below.
3.
Define transient time meaning the time that
system can take to reach a new set point
without triggering the alarm.
4.
The audible alarm is enabled by ticking
“Buzzer”. The alarm is always visual. To
disable alarm remove tick from “Enabled”.
After this device will not alert from deviating
temperatures, only if problems occur.
5.
6.
Press “Save” to save changes.
Alarms log:
By clicking “Log” all led status changes
can be viewed.
This data can be stored to USB drive by
clicking ”To USB” or delete by clicking
“Erase”.
The transient time should be twice the time
that heat plate normally takes to heat to the
set temperature