13
Tighten the filling station to gas cylinder. Make sure it is properly
attached.
Make sure that filling station CONTROL KNOB is turned all the way
counter‐clockwise. Do not use force to operate the knob. Open
gas cylinder slowly and let the pressure flow to the filling station.
BULK gauge shows pressure left in filling gas cylinder.
OUTPUT gauge show pressure that will be dispensed to your gas
bottle.
Dispensed pressure can be adjusted by the output pressure
ADJUSTMENT SCREW. Pressure can be set by turning the screw
with 3/16 inch Allen key. By turning clockwise the output pressure will rise and counter‐clockwise it will
decrease. After you have set the desired pressure (maximum 1500psi), let air out by turning control knob
and you will notice that the output pressure changes from set value. Repeat this pressure adjustment and
air release cycle until the pressure reaches desired value after the air is released. This venting step is
important especially when values are set to lover range. Check the unit ALWAYS before use, because the
values can change over time.
Make sure all the bottle attachment components are unharmed
and air flow from venting hole is not blocked. Attach the bottle to
the filling station by pushing the bottle attachment connector
backwards. After bottle is properly attached and locked on the
connector, the connector should slide back to its place and make
click sound.
Turn and hold the control knob clockwise until gas start to flow to
bottle. After this, turn the control knob ¼ further. Wait for
pressure flowing to stop and let the pressure settle for at least
30seconds. Follow the gas bottle pressure indicator that it won’t
exceed 1500 psi.
When the bottle is full turn the control knob counter‐clockwise to
vent the filling station fully empty.
Detach bottle from regulator and close gas cylinder.
WARNING Always check that OUTPUT pressure does not
exceed 1500 PSI at any circumstances!
WARNING If gas bottle is not detaching easily pressure may still
occupy components! Empty the filling station properly by
turning the control knob.
WARNING! Always operate on completely empty filling station.
Figure 3 – Filling station to gas cylinder
Figure 4 ‐ Filling sttion to bottle
Figure 5 ‐ Control knob opening