26
Ramp rate:
This menu allow user to define maximum
heating rate for the heat plate.
Select the maximum heating rate from slide bar
or + or – and press save.
Settings: Calibrations
This menu calibrates heating device and
temperature sensor.
WARNING If heating ramp rate is defined in high
values, this may lead heating device overshoots
and the temperatures can reach significantly
higher value than defined set value.
User is recommended to calibrate heating
device when first assembled the device and
when the temperature in the room has
changed more than 1 °C.
Temperature sensor is recommended to be
calibrated once a year.