background image

 

Page 23 of 50 

Workstation Exit 

Remove Cap from storage brackets, and ensure the handle is oriented in a vertical position 
on  the  Cap.  The  Handle  is  designed  with  “indexing”  detent  features  to  help  locate  the 
Handle relative to the Cap. 

Drawing  the  handle  into  the  Glove  Port,  replace  the  cap,  using  the  location  posts  and 
graphic to orient the Cap correctly on the Glove Port.  

Note: Care should be taken to ensure that the  Ezee-Sleeve material does not  become 
caught between the Cap and the Glove Port seal.
  

 

Figure 20: Glove port cap 

 

Figure 21: Glove Port Locator 

Rotating the Handle by 90 degrees to the horizontal position will now lock the Cap to the 
Glove Port, allowing withdrawal of the users arm from the Ezee-Sleeve. 

 

Содержание Concept-M

Страница 1: ...Affix Serial Number Sticker Here UserManual UM 039 Concept M Anaerobic and Microaerophilic Workstation Concept M 400 Concept M 500...

Страница 2: ...LING OF THE CONCEPT WORKSTATION 6 ENVIRONMENTAL OPERATING CONDITIONS 6 SERVICE REQUIREMENTS 7 ELECTRICAL SUPPLY REQUIREMENTS 7 GAS SUPPLY REQUIREMENTS 8 CONCEPT OVERVIEW 9 FRONT VIEW 9 FRONT INTERIOR...

Страница 3: ...ELP SETTINGS 33 WARNINGS 34 USB TREND LOG 38 CLEANING AND SERVICE REQUIREMENTS 39 SERVICE AND CLEANING OVERVIEW 39 CLEANING PROCEDURE DURING AND AFTER EACH USE 40 CLEANING PROCEDURE DEEP CLEAN WITH FR...

Страница 4: ...e for Entry 19 Figure 13 Arm inserted into Ezee Sleeve Shown with and without Ezee Sleeve 19 Figure 14 Foot pedals for right and left glove ports 20 Figure 15 Ezee Sleeve before vacuum 20 Figure 16 Ez...

Страница 5: ...30 Screen 10 Utilities Menu 31 Screen 11 Oxygen Sensor Calibration 32 Screen 12 Help Settings Menu 33 Screen 13 Temperature Alarm Screen 34 Screen 14 USB Memory Warning Screen 35 Screen 15 Gas Mix Ti...

Страница 6: ...on Dioxide CO2 and Compressed Air O2 or 25 Oxygen in Nitrogen to achieve the desired atmosphere The user can select the Oxygen concentration from 0 1 to 20 9 and Carbon Dioxide concentrations from 0 1...

Страница 7: ...r Over pressure could damage the workstation Only the gases specified in this user manual may be used All gas bottles must be adequately secured before connection to the workstation The maximum power...

Страница 8: ...ls on standards tested refer to the DoC Certificate WEEE This equipment must be disposed of in accordance with the Waste from Electrical and Electronic Equipment WEEE Directive This product must not b...

Страница 9: ...ference in which case the user will be required to correct the interference at his own expense Precautions The Concept has been tested and approved to EN61010 1 61326 1 and 61326 2 6 This means that t...

Страница 10: ...r to this manual in all cases where this symbol appears in order to find out the nature of the Potential Hazard and actions to be taken in order to avoid the Hazard Warning this equipment contains hig...

Страница 11: ...elocated by a qualified engineer To arrange installation or relocation please contact your local distributor The mains appliance coupler and plug are the AC mains supply isolation device and must be e...

Страница 12: ...voltage and frequency for Concept workstations are Voltage Range Frequency Nominal Power Rated Current Rated Power 220 to 240 Vac 50 60Hz 360W 8 1A 1830W 110 to 120 Vac 50 60Hz 360W 14 3A 1600W 100 V...

Страница 13: ...aximum of 25 Oxygen in Nitrogen Industrial or medical Air For oxygen levels between 17 and 23 42 to 49psi 3 to 3 5 bar Table 5 Additional gases for specific operating modes Figure 2 Gas Connection Gra...

Страница 14: ...ure 3 and Figure 4 respectively Figure 3 Standard Concept Workstation with SPES Option Figure 4 Large Concept Workstation 1 Ezeeyin Glove Ports 2 Single Plate Entry System SPES Optional Accessory 3 Re...

Страница 15: ...ternal View Figure 6 Left Side Internal View 1 Internal Light 2 Temperature and Humidity Sensor Housing 3 Internal interlock door 4 Internal sockets 5 Condensate plate 6 Universal Port Optional 7 Wast...

Страница 16: ...View Figure 7 Left Side End Panel with Ultrasonic Humidity Option 1 Lid Service access only 2 Humidity water tank optional 3 Side Panel Service access only 4 Cooling vent Do not obstruct 5 Universal...

Страница 17: ...Page 12 of 50 Right Side View Figure 8 Right Side End Panel 1 Cooling vent Do not obstruct 2 Side Panel Service access only 1 2...

Страница 18: ...ly Connections Figure 10 Gas and Electrical Connections 1 Safety Label 2 Rear panel Service access only 3 Ethernet Port For use with Remote Monitoring Option Only See Service Requirements section for...

Страница 19: ...nterlock Overview The interlock consists of 4 main components Interlock outer door Interlock inner door Interlock chamber Interlock floor tray The interlock is accessed by opening either the outer or...

Страница 20: ...d closed as often as required to transfer material providing the outer door has not been opened Transferring material into the workstation chamber via the interlock To transfer material into the works...

Страница 21: ...interlock purge cycle has completed indicated by the touchscreen and the inner door push button illumination press the inner door button Slide the interlock inner door backwards to open the interlock...

Страница 22: ...al from the workstation chamber into the interlock chamber Close the interlock inner door by sliding the interlock inner door forwards Exit the workstation chamber via the Ezeeyin glove ports Open the...

Страница 23: ...is provided via the Ezeeyin glove ports and Ezee Sleeve The Ezee Sleeve consists of a gloveless gas tight sleeve and a cuff The Ezee Sleeve attaches to the glove port via two O rings Note The worksta...

Страница 24: ...zee Sleeve before inserting the hand and arm as shown in Figure 12 Figure 12 Compressed Ezee Sleeve for Entry Hold the desired Ezee Sleeve cuff with the opposite hand Support the Ezee Cuff push the ha...

Страница 25: ...ion should be continued until the maximum amount of external atmosphere has been removed from the Ezee Sleeve and the Ezee Sleeve exerts some pressure on the arm hand Figure 15 Ezee Sleeve before vacu...

Страница 26: ...Cap can require a reasonable amount of force This can be made easier by pushing the uppermost part of the Handle to first break the vacuum at the upper edge of the Cap seal Figure 17 Glove port cap T...

Страница 27: ...Page 22 of 50 Figure 19 Glove port cap storage location inside workstation Repeat the procedure for the other hand if both hands are entering the workstation chamber...

Страница 28: ...the Glove Port replace the cap using the location posts and graphic to orient the Cap correctly on the Glove Port Note Care should be taken to ensure that the Ezee Sleeve material does not become caug...

Страница 29: ...orkstation allows the humidity to rise to a user defined maximum at which point it will start to strip the excess humidity away at a control accuracy of 6 One sachet should deliver a humidity level of...

Страница 30: ...below the dew point to allow for the humidity to condense on the plate Figure 24 Condensate Plate As this forms droplets these are then collected into a small reservoir in the end panel that will the...

Страница 31: ...EENS Main screen When the unit is first powered on a Ruskinn Logo screen will appear and will be replaced by Screen 3 Screen 3 Main Menu From this menu you will be able to access all of the features o...

Страница 32: ...ow Alarm Levels Table 6 Settings screen column functions Rows 2 3 and 4 can be changed by the user by pressing on the value and a pop up will appear for the new setting to be selected as shown in Scre...

Страница 33: ...creen will start the gas control Pressing the Home button on the bottom left will return to the main menu If an Oxygen set point of 0 is selected then the settings screen will update to remove the low...

Страница 34: ...use to monitor the current workstation conditions and gain access to all of the in use functions Screen 7 Environment Control Overview Pressing the home icon on the bottom right will bring up Screen 8...

Страница 35: ...wn at the top of the box The red horizontal lines show the current Alarm levels from the settings screen to easily see if there have been any deviations outside of the required operating parameters Pr...

Страница 36: ...e left turns the light on off and the bar will adjust the brightness of the lights in the main work area This bar can either be slid into position or tapping on the setting required Icon 2 Internal Li...

Страница 37: ...ation point and the ambient atmosphere for the high operation point The Atmospheric Oxygen level is set to 20 9 as default but this can be changed if required Once this setting is entered to the corre...

Страница 38: ...nterlock door when the button is pressed for 3 seconds Icon 2 Internal Lights This will open a separate pop up shown in Icon 2 The icon on the left turns the light on off and the bar will adjust the b...

Страница 39: ...ing the initial turn in or if the set point has been changed by more than 0 2 C The alarm can be dismissed by pressing the return arrow at the bottom right of the screen If the workstation still has n...

Страница 40: ...ved for longer than 5 minutes as a reminder to replace it after the log files have been copied to a computer for analysis This alarm can be dismissed by pressing the back arrow at the bottom right of...

Страница 41: ...workstation to avoid excessive gas usage as it indicates an issue with the gas tightness of the workstation Screen 15 Gas Mix Time Limit Warning Screen The alarm can be dismissed by pressing the back...

Страница 42: ...e refer to the section for common problems and solutions Screen 16 Input Gas Pressure Alarm Alarm Purpose Alarm Buzzer Sound Temperature Alarm 1 second Pulse On Off Input Gas Pressure Alarm Continuous...

Страница 43: ...t point CH9 Chamber Temp Actual Chamber Temperature Measurement in C CH10 Water Tank Level For workstations with Ultrasonic Humidity this is the water level remaining in the Water Tank CH11 Interlock...

Страница 44: ...use End User Empty waste port bag Weekly End User Calibrate oxygen sensors Monthly End User Deep clean workstation 3 6 months End User Replace detox sachets Annually End User Qualified service enginee...

Страница 45: ...e permitted The use of UV light is prohibited in the workstation as it will cause permanent damage The use of UV Light within the Workstation will invalidate the warranty Cleaning procedure during and...

Страница 46: ...ble alarm in all screens other than the main menu The following screen will appear to ensure that the glove port pneumatic connector is disconnected before the screen is removed Figure 26 Front Screen...

Страница 47: ...he workstation o Interlock side walls o Interlock ceiling o Interlock floor trays o Interlock outer door For all workstation components wipe with a paper towel soaked in an appropriate cleaning agent...

Страница 48: ...ation chamber Close the outer interlock doors Plug the workstation back in at the mains Switch the workstation on Go into the setting menu reinstall the front screen activate the seal and connect the...

Страница 49: ...ner Sleeves Detox Sachet Service Instructions Follow the Deep Cleaning Preparation stage Unscrew the 2 thumb screws retaining the ceiling panel in the workstation Figure 27 Ceiling Panel Thumb Screws...

Страница 50: ...e 30 Large Catalyst Sachets Replace the ceiling panel and the thumbscrews Remove the used sleeves from the front screen and dispose of them Figure 31 Gloveport Sleeves Follow the Deep Cleaning Cleanin...

Страница 51: ...station to identify low input gas pressures The gas input pressure is set to low o Check the gas supply bottle pressure is set to the recommended levels defined in the Gas Supply Requirements section...

Страница 52: ...n If the problem persists after the above have been carried out then please contact your distributor to assist in identifying the issue or to arrange a service engineer visit Internal Equipment not po...

Страница 53: ...A PARTICULAR PURPOSE THE IMPLIED WARRANTY OF MERCHANTABILITY OR ANY IMPLIED WARRANTY ARISING OUT OF A COURSE OF DEALING OR OF PERFORMANCE CUSTOM OR USAGE OF TRADE Ruskinn Technology Limited DOES NOT...

Страница 54: ...household waste site Instead it should be taken to the appropriate collection point for the recycling of electrical and electronic equipment Alternatively please contact your local distributor for dis...

Страница 55: ...ruskinn com General enquiries ruskinnoffice ruskinn com Website www bakerruskinn com YouTube channel http www youtube com ruskinntechnology Ruskinn Technology Limited is a registered company in United...

Отзывы: