SERVICE REQUIREMENTS
Electrical Supply Requirements
The workstation must be connected to a mains power supply. A power cord is supplied to
connect the workstation to the mains supply. If an alternative power cord is used is must
be rated appropriately for the power requirements of the workstation, refer to Table 2. The
workstation must be connected to a protective earth.
To ensure safe operation of the workstation, it must be connected to a supply of the
correct voltage and frequency as stated on the rating label shown at the rear of the unit.
The mains supply voltage fluctuations must not exceed +/- 10% of the nominal mains
voltage. The input voltage and frequency for Concept workstations are:
Voltage Range
Frequency
Nominal Power Rated Current
Rated Power
220 to 240 Vac
50/60Hz
360W
8.1A
1830W
110 to 120 Vac
50/60Hz
360W
14.3A
1600W
100 Vac
50/60Hz
360W
15.7A
1570W
Table 2: Electrical Service Requirements
Please note that the applicable ratings for the Workstation configuration are detailed the
labelling at the rear of the product as shown in Figure 1.
Figure 1: Concept Rear Connections
1.
Gas Supplies
2.
Residual Current Circuit Breaker (RCCB)
3.
Mains Input Connector (C19)
4.
Footswitch Connector
5.
Mains Input Fuses
The Workstation is double pole fused (Live and Neutral) for protection against over current
draw. The fuses fitted are based on the voltage range applicable, see Table 3 for details:
Voltage Range
Mains Input Fusing
220-240 Vac
T10A H250V
100 Vac, 110-120 Vac
T16A H250V