Chapter 4. Error Codes and Troubleshooting
56
PanaFlow Z3 User’s Manual
S17: Temperature
Temperature error
Error:
Temperature of the electronics is outside the
pre-defined operating range. Make sure that the
ambient temperature is not outside the meter
operating range
S18: Driver Fault
Driver failure
Fault:
Driver failure. Try power cycling the meter. If
error persists after power cycle, contact
BakerHughes factory
S19: Watch Dog Failure
Watch dog failure
Fault:
Watch dog test failed. Try power cycling the
meter. If error persists after power cycle, contact
BakerHughes factory
S21: Stack Overflow
Stack overflow
Fault:
Stack overflow. Try power cycling the meter. If
error persists after power cycle, contact
BakerHughes factory
S22: Sequence or Window
Watchdog
Sequence failed
Fault:
Sequence failure detected. Try power cycling
the meter. If error persists after power cycle, contact
BakerHughes factory
S23: Initialization Failed
Initialization failed
Error:
Initialization failed. Please verify all the
configuration parameters. If error persists, contact
BakerHughes factory
S24: DSP Hardware Errors
DSP hardware failed
Fault:
DSP hardware failure detected. Try power
cycling the meter. If error persists after power cycle,
contact BakerHughes factory
S25: DSP Exception
DSP exception
Fault:
DSP exception. Try power cycling the meter. If
error persists after power cycle, contact
BakerHughes factory
S26: Default ISR
Exception within the ISR
Fault:
Exception within the ISR. Try power cycling the
meter. If error persists after power cycle, contact
BakerHughes factory
S27: DSP Reset ISR
Exception within the DSP ISR
Fault:
Exception within the DSP ISR. Try power cycling
the meter. If error persists after power cycle, contact
BakerHughes factory
S28: Software Fault
Software malfunction
Error:
Software malfunction. Try power cycling the
meter. If error persists after power cycle, contact
BakerHughes factory
S30: Flash Save Failed
Save to Flash Failed
Error:
Request to Save failed. Try again. If error
persists, contact BakerHughes factory.
Table 20: System Error Description and Recommended Actions
Error Code
Error Message
Description / Recommended Action
Содержание Panametrics PanaFlow Z3
Страница 1: ...Flow 910 311 Rev C April 2019 Panametrics com PanaFlow Z3 User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 8: ...Contents vi PanaFlow Z3 User s Manual...
Страница 12: ...Preface x PanaFlow Z3 User s Manual no content intended for this page...
Страница 21: ...PanaFlow Z3 User s Manual 9 Chapter 2 Installation Figure 8 Lifting the PanaFlow Z3 System...
Страница 25: ...PanaFlow Z3 User s Manual 13 Chapter 2 Installation Figure 12 Remote Mount Wiring Diagram...
Страница 26: ...Chapter 2 Installation 14 PanaFlow Z3 User s Manual no content intended for this page...
Страница 58: ...Chapter 3 Programming 46 PanaFlow Z3 User s Manual no content intended for this page...
Страница 74: ...Chapter 4 Error Codes and Troubleshooting 62 PanaFlow Z3 User s Manual no content intended for this page...
Страница 85: ...PanaFlow Z3 User s Manual 73 Appendix A Specifications and Model Configurations A 4 Serial Number Information...
Страница 86: ...Appendix A Specifications and Model Configurations 74 PanaFlow Z3 User s Manual...
Страница 104: ...Appendix B Data Records 92 PanaFlow Z3 User s Manual no content intended for this page...
Страница 105: ...PanaFlow Z3 User s Manual 93 Appendix C Modbus Map Appendix C Modbus Map C 1 Input Registers Map...
Страница 108: ...Appendix D CE Mark Compliance 96 PanaFlow Z3 User s Manual no content intended for this page...
Страница 110: ...Warranty 98 PanaFlow Z3 User s Manual no content intended for this page...
Страница 115: ...133M6993 Rev B 103 of 104 Certification Safety Statements for PanaFlow PF10 Zx Apr 2019 LOCAL MOUNT WIRING...
Страница 116: ......
Страница 117: ......