Chapter 4. Error Codes and Troubleshooting
48
PanaFlow Z3 User’s Manual
2.
Defective/Damaged cables, transducers, couplant, buffer or electronics.
After you have tried eliminating/correcting for any most likely causes mentioned above, if error still exists, also check
Process/flow conditions change.
1.
Excessive turbulence.
2.
Discontinuities in fluid characteristics such as multi-phase flow, flashing, pockets of gas, presence of bubbles or
solid particles, cavitation or rapidly changing fluid type.
3.
Extreme fluid properties, such as pressure or temperature.
4.
Wax build-up on face of buffer/transducer. Clogged ports due to sedimentation.
5.
Half-full pipe.
4.3.1.2
Multi-Channel Error
If more than one channel is in error, the most likely cause is changes in process/flow conditions.
1.
Excessive turbulence.
2.
Discontinuities in fluid characteristics such as multi-phase flow, flashing, pockets of gas, presence of bubbles or
solid particles, cavitation or rapidly changing fluid type.
3.
Extreme fluid properties, such as pressure or temperature.
4.
Wax build-up on face of buffer/transducer. Clogged ports due to sedimentation.
5.
Half-full pipe.
After you have tried eliminating/correcting for any most likely causes mentioned above, if error still exists, also check,
1.
Incorrect programming on Error Limits or flow condition changes that now make previous programming invalid.
2.
Defective/Damaged cables, transducers, couplant, buffer or electronics.
In case you are unable to clear the errors, collect diagnostic data and parameter information for each channel in the
Diagnostics Data Table
prior to calling your local sales or service center.
4.3.1.3
Viewing Channel Specific Error/Warnings
To indicate the health of the meter, PanaFlow Z3 has built-in Error codes. The Channel specific errors are very critical
in determining the corrective actions required Figure 40 shows the steps to view current channel specific
errors/warnings. The description of the Error Codes and the recommended actions are provided in Table 17.
Figure 40: Viewing Current Chanel Specific Errors
4.3.1.4
Viewing Flow diagnostics
To enable troubleshooting with additional flow diagnostics PanaFlow Z3 has Flow Info page that allows the user to
scroll through all critical flow diagnostics. The Channel specific errors are very critical in determining the corrective
actions required. Figure 40 shows the steps to view current channel specific errors/warnings. The description of the
Error Codes and the recommended actions are provided in Table 17.
Main Menu
Errors/Warning
9
2
Errors/Warning 1/1
9
E23: MultiChAccuracy
2
Velocity
0.000
m/s
Display Format
Program
Flow Info
CH1 E1: SNR Error!
CH2 E1: SNR Error!
CH3 E1: SNR Error!
Содержание Panametrics PanaFlow Z3
Страница 1: ...Flow 910 311 Rev C April 2019 Panametrics com PanaFlow Z3 User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 8: ...Contents vi PanaFlow Z3 User s Manual...
Страница 12: ...Preface x PanaFlow Z3 User s Manual no content intended for this page...
Страница 21: ...PanaFlow Z3 User s Manual 9 Chapter 2 Installation Figure 8 Lifting the PanaFlow Z3 System...
Страница 25: ...PanaFlow Z3 User s Manual 13 Chapter 2 Installation Figure 12 Remote Mount Wiring Diagram...
Страница 26: ...Chapter 2 Installation 14 PanaFlow Z3 User s Manual no content intended for this page...
Страница 58: ...Chapter 3 Programming 46 PanaFlow Z3 User s Manual no content intended for this page...
Страница 74: ...Chapter 4 Error Codes and Troubleshooting 62 PanaFlow Z3 User s Manual no content intended for this page...
Страница 85: ...PanaFlow Z3 User s Manual 73 Appendix A Specifications and Model Configurations A 4 Serial Number Information...
Страница 86: ...Appendix A Specifications and Model Configurations 74 PanaFlow Z3 User s Manual...
Страница 104: ...Appendix B Data Records 92 PanaFlow Z3 User s Manual no content intended for this page...
Страница 105: ...PanaFlow Z3 User s Manual 93 Appendix C Modbus Map Appendix C Modbus Map C 1 Input Registers Map...
Страница 108: ...Appendix D CE Mark Compliance 96 PanaFlow Z3 User s Manual no content intended for this page...
Страница 110: ...Warranty 98 PanaFlow Z3 User s Manual no content intended for this page...
Страница 115: ...133M6993 Rev B 103 of 104 Certification Safety Statements for PanaFlow PF10 Zx Apr 2019 LOCAL MOUNT WIRING...
Страница 116: ......
Страница 117: ......