PanaFlow Z3 User’s Manual
41
Chapter 3. Programming
3.6
Programming Menu Options
The options in the Programming Page should be selected to best suit your application. The configurations selected in
programming page are critical for accurate flow measurements. Incorrect programming settings can give
erroneous measurements and impact accuracy.
Note:
Consult the factory or BakerHughes Services if you are unsure of the appropriate settings for your
application.
IMPORTANT: The configurations in the Programming page are pre-selected to best suit your applications. Consult
the factory or BakerHughes Services before changing any of these settings. Modifying any of the
settings can give erroneous measurements and impact accuracy.
3.6.1
Programming the Pipe
The Pipe menu allows the user to specify all pipe parameters that are required to ensure accurate ultrasonic flow
rate measurements. Use steps as in Section 3.3.2 to navigate to the Programming page.
1.
Highlight
[Composite]
and press
[ENTER]
. Then Select
[Pipe]
and press
[ENTER]
.
2.
Pipe dimensions like
[Outer Diameter]
(OD),
[Wall Thickness]
and
[Inner Diameter]
(ID) can be programmed
in this menu.
Note:
The measurement units used for pipe size parameters depend on the choices made in the Section 3.4.1.
Figure 34: Pipe Programming
3.6.2
Programming the Fluid
The
Fluid
menu (see Figure 35) allows the user to specify all the parameters of the fluid flowing through the pipe that
is required to ensure accurate ultrasonic flow rate measurements. Use steps as in section Section 3.3.2 to navigate to
the Programming page.
1.
Select
[Composite]
and press
[ENTER].
Scroll down, select
[Fluid]
and press
[ENTER]
.
2.
Select
[Density]
, press
[ENTER]
and program the actual density
[Density (Act)]
and reference density
[Density
(Ref)]
of the process fluid.
3.
Then select
[Kinematic Viscosity]
, press
[ENTER]
and program the kinematic viscosity of the process fluid.
4.
Then select
[Tracking]
option. The tracking window is used to scan through the speed of sound range
programmed to detect the signal when the user is unsure of the fluid sound speed. The meter also supports a list
of standard fluid types. If the process fluid is not listed in standard fluids list and if you are unsure of the fluid
sound speed, set tracking window to On and program the
minimum
and
maximum
sound speed range to scan.
Tracking window should be set to On.
5.
For standard fluids list (see Table 15) supported in the meter, the
minimum
,
maximum
and
nominal sound
speeds
are automatically selected.
6.
Also program the process
[Fluid Temperature]
and
[Ambient Temperature]
IMPORTANT: Consult the factory before setting Tracking Window to Off.
Programming
Composite
Prigram/Composite
Path
Fluid
Pipe
Limits
Prigram/Composite
Fluid
Channel 1
Pipe OD
UNIT: In
3.50
Channel 2
Advanced
Channel 3
Pipe
Pipe Size
Pipe Wall Thickness
Pipe OD
Path
Limits
Pipe Size
Pipe ID
9
2
9
2
9
2
or
9
2
9
Pipe OD
3.50 in
Pipe OD
9
2
Operator Access or Higher
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
Содержание Panametrics PanaFlow Z3
Страница 1: ...Flow 910 311 Rev C April 2019 Panametrics com PanaFlow Z3 User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 8: ...Contents vi PanaFlow Z3 User s Manual...
Страница 12: ...Preface x PanaFlow Z3 User s Manual no content intended for this page...
Страница 21: ...PanaFlow Z3 User s Manual 9 Chapter 2 Installation Figure 8 Lifting the PanaFlow Z3 System...
Страница 25: ...PanaFlow Z3 User s Manual 13 Chapter 2 Installation Figure 12 Remote Mount Wiring Diagram...
Страница 26: ...Chapter 2 Installation 14 PanaFlow Z3 User s Manual no content intended for this page...
Страница 58: ...Chapter 3 Programming 46 PanaFlow Z3 User s Manual no content intended for this page...
Страница 74: ...Chapter 4 Error Codes and Troubleshooting 62 PanaFlow Z3 User s Manual no content intended for this page...
Страница 85: ...PanaFlow Z3 User s Manual 73 Appendix A Specifications and Model Configurations A 4 Serial Number Information...
Страница 86: ...Appendix A Specifications and Model Configurations 74 PanaFlow Z3 User s Manual...
Страница 104: ...Appendix B Data Records 92 PanaFlow Z3 User s Manual no content intended for this page...
Страница 105: ...PanaFlow Z3 User s Manual 93 Appendix C Modbus Map Appendix C Modbus Map C 1 Input Registers Map...
Страница 108: ...Appendix D CE Mark Compliance 96 PanaFlow Z3 User s Manual no content intended for this page...
Страница 110: ...Warranty 98 PanaFlow Z3 User s Manual no content intended for this page...
Страница 115: ...133M6993 Rev B 103 of 104 Certification Safety Statements for PanaFlow PF10 Zx Apr 2019 LOCAL MOUNT WIRING...
Страница 116: ......
Страница 117: ......