Copyright 2003 Baker Hughes Company.
English–DPI800/802 Instruction Manual | 5
Example 2) Set up values for x:y Scaling = 0 to 100%
2. Operation
This section gives examples of how to connect and use the
instrument. Before you start:
•
Read and understand section “Safety” on page
•
Do not use a damaged instrument.
2.1 Pressure Connections
Use an applicable method to seal the pressure
connections, and then tighten to the applicable torque
1).
Note:
If the instrument uses G1/8 connections, and the
pressure is ≥
1500
psi (100
bar), connection method (b) is
mandatory.
Figure
1: Connection Methods
2.2 Communications Port Connections
Use the communications port (Figure
item [9]) to
attach an IDOS Universal Measurement Module (UMM).
When you attach the cable from a UMM (Figure
8), the
instrument automatically changes the menus to give you all
the applicable options (Table
3).
2.3 Measure Pressure
To measure the pressure or the pressure difference (if
there are two IDOS pressure ranges).
1. Connect the instrument (Figure
2) and, if necessary,
adjust the Set Up (Table
1).
2. Select an applicable pressure task from Select Task
3) and, if necessary, adjust the
Settings (Table
4).
3. If necessary, do a zero correction (Table
4).
4. Before you disconnect the instrument, isolate and
bleed the system.
Figure
2: Example Configuration - Measure Pressure
The examples in Figure
2 shows:
•
How to measure one or two pressures in a system.
•
How to measure pressure difference.
Note:
∆P = High range IDOS - Low range IDOS. If
Pressure
(p2)
>
Pressure
(p1), the result is negative.
2.4 Leak Test
To do a leak test on a pressure system:
1. Connect the instrument (Figure
3) and, if necessary,
adjust the Set Up (Table
1).
2. Select an applicable leak test from Select Task
3) and, if necessary, adjust the
Settings (Table
4).
3. Set the period for the leak test (Table
4).
4. If necessary, do a zero correction (Table
4).
5. To start the leak test, select Start (
). When the test
is finished, the instrument calculates the leak rate in the
applicable units/minute.
Settings
x:y
1
Settings
x:y
2
Edit
3
4
+
Sk1 = Edit
= Flow scaling (mA, pressure only)
CAUTION
To prevent damage, do not apply
torque to the body of the instrument. If
available, use the flat faces on the pressure
connector to hold the instrument in position.
a 1/8 NPT connections.
b G1/8 connections: Pressure ≥ 1500 psi (100 bar)
c G1/8 connections: Pressure < 1500 psi (100 bar)
d M5 or 10
-
32 UNF connections: Refer to Table
9.
1 Process connector. Maximum torque:
1/8 NPT: 26 lbf.ft (35 Nm)
G1/8: 18.4 lbf.ft (25 Nm)
2 Applicable instrument pressure connector (Table
9).
3 (1/8 NPT only) Thread with an applicable sealant.
4 (G1/8 only) Applicable bonded seal.
5 Reference port connector: M5 or 10
-
32 UNF (Table
Maximum torque: 1.5 lbf.ft (2 Nm)
Sk1
=
=
%
mA
+04.000
+20.000
+04.000
+20.000
=
+20.000
=
+04.000
=
+100.00
=
+00.000
MENU
OK
1
2
3
1
2
4
1
2
4
5
2
a
b
c
d
P
MENU
OK
HOLD
ESC
P
5.003
P1-P2
MENU
OK
'
'
P1 ... g
P1 30 ... g
15.006
30 ... g
MENU
OK
10.003
20 ... g
p1
p2
P1
P2
p1
p2
Settings
Settings
a) P1 and/or P2
b) P1 - P2
Содержание Druck Sensycal DPI800
Страница 2: ......
Страница 36: ...Copyright 2003 Baker Hughes Company 32 DPI800 802 Bedienungsanleitung Deutsch...
Страница 48: ...Copyright 2003 Baker Hughes Company 44 Manual de instrucciones DPI800 802 Espa ol...
Страница 60: ...Copyright 2003 Baker Hughes Company 56 DPI800 802 Notice d emploi Fran ais...
Страница 72: ...Copyright 2003 Baker Hughes Company 68 Manuale d uso DPI800 802 Italiano...
Страница 84: ...Copyright 2003 Baker Hughes Company 80 Manual de Instru es do DPI800 802 Portugu s...
Страница 119: ......