5.3 Electrical Connectors
Refer to Figure
Table
11: Maximum Pressure (Reference Port)
Ranges: g/d Only
MWP
≤ 5 psi (350 mbar)
2 x FS
10 to 15 psi (700 mbar to 1 bar)
1.2 x FS
≥ 30 psi (2 bar)
30 psi (2 bar)
Item
Specification
Range (Measure)
0 to ±55
mA
Accuracy
0.02% of r 3 counts
Temperature coefficient
14 … 50°F, 86 … 122°F
(
-
10 … 10°C, 30 … 50°C)
0.0011% FS / °F
(0.002% FS / °C)
Switch detection
Open and closed. 2
mA current.
Loop power output
24 V ± 10%
HART® resistor
250
Ω (menu selection)
Connectors (Figure
A2)
Three 0.16” (4
mm) sockets