4. Start the test. When it finishes, compare the result with the specified leak rate. Refer to
7.5.2.2 Maximum Vacuum Test
1. PV
621G / PV
621
-
IS / PV
622G / PV
622
-
IS models only. Use the applicable procedure for
the pressure station to set the maximum vacuum:
•
PV
621G / PV
621
-
IS models: Chapter
3 Set the pressure to
-
950
mbar.
•
PV
622G / PV
622
-
IS models: Chapter
4 Set the pressure to
-
950
mbar.
2. Set the DPI
620G / DPI
620G
-
IS calibrator to do a Leak Test:
•
Channel Settings: Units = Bar; Utility = Leak Test
•
Settings: Test Time = 00:01:00 (1
minute)
3. Let the pressure become stable for 1
minute.
4. Start the test. When it finishes, compare the result with the specified leak rate. Refer to
7.6 Return Goods/Material Procedure
If the unit requires calibration or is defective, return it to the nearest Druck Service Centre listed
at:
Contact the Service Department to obtain a Return Goods/Material Authorization (RGA or RMA).
Provide the following information for a RGA or RMA:
•
Product (e.g. PV
621G)
•
Serial number.
•
Details of defect/work to be undertaken.
•
Calibration traceability requirements.
•
Operating conditions.
8. Specification
For a full specification of the PV
62XG / PV
62X
-
IS
pressure stations, refer to the datasheet.
CAUTION
PV
623G / PV
623
-
IS models only. Ice in the pressure mechanism
can cause damage. If the temperature is less than 4°C (39°F), drain all water from
the pressure station.
Table
1: General Specification
Item
Description
Operating temperature
-
10 to 50°C (14 to 122°F)
PV
623G / PV
623
-
IS models only: See caution.
Storage temperature
-
20 to 70°C (
-
4 to 158°F)
PV
623G / PV
623
-
IS models only: See caution.
Pressure safety
Pressure Equipment Directive - Class: Sound Engineering Practice (SEP)