English–PV 62X / PV 62X-IS Instruction Manual | 13

7.3  Cleaning

Clean the case with a moist, lint-free cloth and a weak detergent. Do not use solvents or abrasive 

materials.

7.4  Draining the Unit (PV

 

623G / PV

 

623

-

IS Models)

7.5  Leak Test

Table

 

2 on page

 

15 gives the maximum leak rates for each type of PV

 

62XG / PV

 

62X

-

IS 

pressure station.

7.5.1  Preparation

To do a leak test, we recommend these items:
1. A DPI

 

620G / DPI

 

620G

-

IS calibrator.

2. The applicable PM

 

620 / PM

 

620

-

IS module for the pressure station:

PV

 

621G / PV

 

621

-

IS models: P/N PM620

-

13G / PM620S

-

13G (20

 

bar)

PV

 

622G / PV

 

622

-

IS models: P/N PM620

-

165G / PM620S

-

165G (100

 

bar)

PV

 

623G / PV

 

623

-

IS models: P/N PM620

-

23A / PM620S

-

23A (1000

 

bar)

3. An applicable blanking adaptor to seal the test port connection.
4. PV

 

623G / PV

 

623

-

IS models only: Demineralized water.

7.5.2  Procedure

1. Seal the test port connection with the blanking plug.
2. Connect the applicable PM

 

620 / PM

 

620

-

IS module:

PV

 

621G / PV

 

621

-

IS models: P/N PM620

-

13G / PM620S

-

13G (20

 

bar)

PV

 

622G / PV

 

622

-

IS models: P/N PM620

-

165G / PM620S

-

165G (100

 

bar)

PV

 

623G / PV

 

623

-

IS models: P/N PM620

-

23A / PM620

-

23A (1000

 

bar)

3. Connect a DPI

 

620G / DPI

 

620G

-

IS calibrator and set the power on.

4. Do the test for maximum pressure or maximum vacuum.

7.5.2.1  Maximum Pressure Test

1. Use the applicable procedure for the pressure station to set the maximum pressure:

PV

 

621G / PV

 

621

-

IS models: Chapter

 

Set the pressure to 20

 

bar.

PV

 

622G / PV

 

622

-

IS models: Chapter

 

Set the pressure to 100

 

bar.

PV

 

623G / PV

 

623

-

IS models: Chapter

 

5 Set the pressure to 1000

 

bar.

2. Set the DPI

 

620G / DPI

 

620G

-

IS calibrator to do a Leak Test:

Channel Settings: Units = Bar; Utility = Leak Test

Settings: Test Time = 00:01:00 (1 minute)

3. Let the pressure become stable for 1

 

minute.

CAUTION  

To prevent damage to the pressure station, do not let dirt get into 

the pressure mechanism. Before you connect equipment, make sure it is clean.

CAUTION  

Ice in the pressure mechanism can cause damage. If the 

temperature is less than 4°C (39°F), drain all water from the pressure station. 

Refer to Section

 

5.7

Содержание Druck PV 62X

Страница 1: ...Druck com PV 62X PV 62X IS Pneumatic Hydraulic Pressure Stations Instruction Manual...

Страница 2: ......

Страница 3: ...ECEx and ETL approved intrinsically safe versions are available for use in hazardous areas Summary of Functions This list gives a summary of the available pressure station functions Quick fit pressure...

Страница 4: ...safety procedures this publication and the instructions for the accessories options and equipment you are using it with Read more safety conditions in the 132M9252 safety guide Model Pressure Range P...

Страница 5: ...PV 622G PV 622 IS 5 4 1 Introduction 5 4 2 Release the Pressure 5 4 3 Operation 5 4 3 1 Vacuum Procedure 5 4 3 2 Pressure Procedure 6 5 Hydraulic Pressure Operation PV 623G PV 623 IS 6 5 1 Introductio...

Страница 6: ...nstruction Manual English 7 Maintenance 12 7 1 Introduction 12 7 2 Repair 12 7 3 Cleaning 13 7 4 Draining the Unit PV 623G PV 623 IS Models 13 7 5 Leak Test 13 7 5 1 Preparation 13 7 5 2 Procedure 13...

Страница 7: ...Chapter 2 3 Pressure and electrical connections for a PM 620 PM 620 IS PM 620T PM 620T IS module If a PM 620 PM 620 IS PM 620T PM 620T IS is not fitted use Druck blanking plug IO620 BLANK 4 Pneumatic...

Страница 8: ...pressure ii Refill Close the refill valve 5 or 10 to refill the pressure mechanism Turn the volume adjuster fully counter clockwise Then turn the volume adjuster clockwise to refill the pressure mecha...

Страница 9: ...sconnect the equipment under test from the pressure station test port 1 Safely release all pressure in the system See Section 3 2 for PV 621G PV 621 IS See Section 4 2 for PV 622G PV 622 IS See Sectio...

Страница 10: ...y counter clockwise 2 To do equal adjustments up or down at the end of the procedure turn the volume adjuster to the middle of its range of operation Note To get the maximum vacuum turn the volume adj...

Страница 11: ...tion of the pressure station controls 4 2 Release the Pressure To release all pressure in the system 1 Open the refill valve counter clockwise 1 turn 2 Open the pressure release valve counter clockwis...

Страница 12: ...el with the handle folded in 7 If the volume adjuster reaches the end of its travel close the refill valve clockwise Note Whilst the refill valve is closed there is no change in pressure to the equipm...

Страница 13: ...ollowing procedure describes how to fill the pressure station with hydraulic fluid and how to prime the system If priming a long hose keep the open end of the hose vertical to reduce trapped air 5 3 1...

Страница 14: ...not put too much fluid into the reservoir Leave a small air gap at the top of the reservoir Note To prevent contamination use only one type of hydraulic fluid in the pressure station 5 Replace the pr...

Страница 15: ...der test or the PM 620 PM 620 IS PM 620T PM 620T IS module 4 Open the refill valve 1 turn 5 Use the volume adjuster to adjust the pressure decrease increase Note At high pressures it is easier to turn...

Страница 16: ...and eye protection A sufficiently large container to hold the hydraulic fluid Applicable materials to clean the pressure station refer to Chapter 7 5 6 2 Procedure 1 Release the pressure Section 5 2 2...

Страница 17: ...m and repeat steps 3 and 4 Note To prevent contamination use only one type of hydraulic fluid in the pressure station 6 Pressure Relief Valve PRV 6 1 Introduction Use a pressure relief valve PRV to li...

Страница 18: ...cable pressure indicator to the test port or use a DPI 620G DPI 620G IS calibrator with a PM 620 PM 620 IS PM 620T PM 620T IS module 2 Remove the plastic cap from the end of the PRV 3 Set the necessar...

Страница 19: ...PM 620 PM 620 IS module PV 621G PV 621 IS models P N PM620 13G PM620S 13G 20 bar PV 622G PV 622 IS models P N PM620 165G PM620S 165G 100 bar PV 623G PV 623 IS models P N PM620 23A PM620 23A 1000 bar...

Страница 20: ...Goods Material Procedure If the unit requires calibration or is defective return it to the nearest Druck Service Centre listed at Druck com Contact the Service Department to obtain a Return Goods Mate...

Страница 21: ...100 bar 13 5 to 1500 psi 0 to 1000 bar 0 to 15000 psi Minimum resolution with a typical test volume 0 001 bar 0 0145 psi 0 005 bar 0 0725 psi 0 1 bar 1 45 psi Pressure system volume 1 V1 Volume adjus...

Страница 22: ...20 PRV P5 Pneumatic 100 bar 1500 psi 60 to 100 bar 870 to 1 500 psi IO620 PRV P6 Pneumatic 3 bar 45 psi 1 1 to 3 bar 16 to 45 psi IO620 PRV P7 Pneumatic 12 bar 170 psi 6 1 to 12 bar 90 to 170 psi IO62...

Страница 23: ......

Страница 24: ...kerhughes com Bangalore 91 9986024426 aneesh madhav bakerhughes com Milan 39 02 36 04 28 42 csd italia bakerhughes com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken 3...

Отзывы: