ii | DPI 620G-IS Safety and Quick Start Guide-English
© 2016 Baker Hughes, a GE company – All rights reserved.
Special Conditions for Safe Use
1.
The DPI 620G-IS USB Client Connection must only be connected to external
apparatus within a Safe Area with U
m
= 254 V.
2.
When a remote sensor is powered using the Channel 2, 15V, U
o
loop supply, the
remote sensor must be disconnected from all other sources of power.
3.
The Channel 2, V terminals can be connected to a FISCO System if the power for the
system is provided from the DPI 620G-IS Channel 2, 15V, U
o
loop supply and the
electrical parameters of the field devices are compatible with those of the
DPI 620G-IS.
4.
The DPI 620G-IS should not be connected to a powered FISCO system unless its
defined electrical parameters are compatible with the DPI 620G-IS.
5.
If the Channel 1 Terminals and the Channel 2 Terminals are used together then they
must always be connected as separate Intrinsically Safe Circuits.
6.
The rechargeable battery pack may be removed from or replaced in a DPI 620G-IS
within a hazardous area.
7.
The rechargeable battery pack must be removed from the DPI 620G-IS for recharging
in the safe area using only the Socket 3 charging contacts and the Druck charger with
U
m
= 254 V. No connections must be made to the IS outputs.
8.
Both the MC 620-IS Dual Transducer Carrier, or the PV 62X-IS Series Pressure Station
have a location for either one or two transducers which screw into the pressure
connection and the two slip rings on the lower face of a transducer make contact with
two spring loaded pins. When either a MC 620-IS Dual Transducer Carrier, or the
PV 62X-IS Series Pressure Station is connected to a DPI 620G-IS Advanced Modular
Calibrator, these pins are energised and do not meet the requirements of at least
IP20. Therefore it is a requirement of certification that before either the MC 620-IS
Dual Transducer Carrier or the PV 62X-IS Series Pressure Station is connected to the
DPI 620G-IS Advanced Modular Calibrator, a PM 620-IS Transducer or a metallic
Dummy Pressure Transducer is screwed into any vacant positions and remains in
position until the assembly is removed from the DPI 620G-IS Advanced Modular
Calibrator.
IECEx Approvals
For the IECEx certificate (IECEx BAS16.0010X), visit the IECEx website at:
INFORMATION
The MC 620-IS Dual Transducer Carrier has ATEX and
IECEx approval that is “Part of” the DPI 620-IS approvals.
(Baseefa10ATEX0010X / IECEx BAS 10.0002X).
The PV 62X-IS Series Pressure Station has separate ATEX and IECEx
approval (Baseefa10ATEX0011X / IECEx BAS 10.0003X).
Содержание Druck DPI 620 Genii
Страница 2: ......
Страница 10: ...viii DPI 620G IS Safety and Quick Start Guide English 2016 Baker Hughes a GE company All rights reserved...
Страница 106: ...viii DPI 620G IS Baker Hughes a GE company 2016...
Страница 109: ...Baker Hughes a GE company 2016 DPI 620G IS 3 1 4 1 4 1 1 5 1 2 2 3 3 4 8 4 3 1 2 4...
Страница 110: ...1 4 DPI 620G IS Baker Hughes a GE company 2016 1 4 2 1 5 1 1 2 2 3 1 2 3...
Страница 112: ...1 6 DPI 620G IS Baker Hughes a GE company 2016 1 8 Druck DPI 620G IS HART Foundation Fieldbus Profibus 1 8 1...
Страница 113: ...Baker Hughes a GE company 2016 DPI 620G IS 7 1 8 2 Settings Date Time Language 1 8 3 DPI 620G IS Help 1 8 4 1 8 4 1...
Страница 114: ...1 8 DPI 620G IS Baker Hughes a GE company 2016 1 9 IEC 61010 1 CAT I CAT IV 1 10 1 11 CAT I CAT I CAT I CAT II...
Страница 122: ...viii DPI 620G IS 2016 Baker Hughes a GE company...
Страница 125: ...2016 Baker Hughes a GE company DPI 620G IS 3 1 4 2 1 5 1 1 2 2 3 1 2 3...
Страница 128: ...1 6 DPI 620G IS 2016 Baker Hughes a GE company 1 8 4 1 8 4 1 1 9 IEC 61010 1 CAT I CAT IV CAT I CAT I CAT I CAT II...
Страница 129: ...2016 Baker Hughes a GE company DPI 620G IS 7 1 10 1 11...
Страница 130: ...1 8 DPI 620G IS 2016 Baker Hughes a GE company...
Страница 138: ...viii DPI 620G IS 2016 Baker Hughes a GE company...
Страница 141: ...2016 Baker Hughes a GE company DPI 620G IS 3 1 4 1 4 1 1 DC 5V DC 1 2 2 3 3 4 8 LED 4 3 1 2 4...
Страница 142: ...1 4 DPI 620G IS 2016 Baker Hughes a GE company 1 4 2 1 5 1 1 2 2 3 1 2 3...
Страница 143: ...2016 Baker Hughes a GE company DPI 620G IS 5 1 6 1 6 1 1 1 6 2 1 1 6 3 1 1 6 4 1...
Страница 146: ...1 8 DPI 620G IS 2016 Baker Hughes a GE company 1 9 IEC 61010 1 CAT I CAT IV 1 10 1 11 CAT I CAT I CAT I CAT II...
Страница 170: ...viii DPI 620G IS Sikkerheds og startvejledning Dansk 2016 Baker Hughes a GE company Alle rettigheder forbeholdes...
Страница 186: ...viii DPI 620G IS biztons gi s r vid tmutat magyar 2016 Baker Hughes a GE company Minden jog fenntartva...
Страница 218: ...viii DPI 620G IS dro bas un s pam c bas rokasgr mata Latvie u 2016 Baker Hughes a GE company Visas ties bas patur tas...
Страница 234: ...viii DPI 620G IS s kerhets och snabbstartguide Svenska 2016 Baker Hughes a GE company med ensamr tt...
Страница 266: ...viii DPI 620G IS Sikkerhets og hurtigstartveiledning Norsk 2016 Baker Hughes a GE company Med enerett...
Страница 298: ...viii DPI 620G IS G venlik ve H zl Ba lang K lavuzu T rk e 2016 Baker Hughes a GE company T m haklar sakl d r...
Страница 314: ...viii Ghid de siguran i de pornire rapid DPI 620G IS Rom n 2016 Baker Hughes a GE company Toate drepturile rezervate...
Страница 330: ...viii DPI 620G IS 2016 Baker Hughes a GE company...
Страница 333: ...2016 Baker Hughes a GE company DPI 620G IS 3 3 4 8 LED 4 1 4 2 1 5 1 1 2 2 3 1 2 3...
Страница 334: ...1 4 DPI 620G IS 2016 Baker Hughes a GE company 1 6 1 6 1 1 1 6 2 1 1 6 3 1 1 6 4 1 7 GE Dark GE Light 2 Settings 1...
Страница 336: ...1 6 DPI 620G IS 2016 Baker Hughes a GE company 1 8 3 DPI 620G IS Help 1 8 4 1 8 4 1...
Страница 337: ...2016 Baker Hughes a GE company DPI 620G IS 7 1 9 IEC 61010 1 4 CAT I CAT IV 1 10 1 11 CAT I CAT I CAT I CAT II...
Страница 338: ...1 8 DPI 620G IS 2016 Baker Hughes a GE company...
Страница 339: ......