Standards and Certifications
STANDAODS AND CEOTIFICATIONS
Safety
Agency Approval/Standards
• CE Marked for Low Voltage Directive and RoHS
• CSA Marked per Class C225286 and C225206, Process Control Equipment
• CSA C22�2 No� 61010-1-12, General requirements
• CAN/CSA-C22�2 No� 61010-1-12 Safety requirements for electrical equipment
for measurement, control and laboratory use� Part 1: General requirements—
Tri-national standard with UL 61010-1 and ANSI/ISA-61010-1 (82�02�01)
EMI/EMC Compliance
Conducted and Radiated
Emissions per CISPR11:2009 /
EN55011�
Class A, Group 1
IEC 61000-4-2:2008
Electro-static Discharge
2/4kV - Contact Discharge,
2/4/8kV Air Discharge Performance
Criteria B
IEC 61000-4-3:2006
Radiated RF Immunity
Test levels: 80-1000 MHz & 1400-2000 MHz
Performance Criteria A
IEC 61000-4-4:2004 EFT Immunity
(Signal and Power lines)
Tested per specification to Performance
Criteria B
IEC 61000-4-5:2005
Surge Protection
Tested per specification to Performance
Criteria B
IEC 61000-4-6:2008
Conducted RF Immunity
(Signal and Power lines)
Test Levels: 0�15-80 MHz
Level 3, Performance Criteria A
IEC 61000-4-11:2004
Voltage Dips, Interruptions,
and Dropouts
Tested per specification to Performance
Criteria B & C
Enclosure Protection
• IEC/CSA/UL 60529-1: Degrees of protection provided by enclosures (IP65),
when installed with all four mounting clips in a similarly rated enclosure, which
includes the optional wall mount enclosure�
• Additional Protection (optional): NEMA 4X (wall mount enclosure only)�
Page 43
July 2016
CTL-UM-01723-EN-01