ID
Parameter
Value
Notes
4 7
To activate energy pulses
output (when volume pulse
output is active)
TEST
tESt on
Wh
Activated by pressing and holding the
button
To activate volume pulse
output (when energy pulse
output is active)
TEST m³
tESt on
Activated by pressing and holding the
button
4 8
To deactivate the
Test Mode
TEST
tESt OFF
Deactivated by pressing and holding the
button
4 9
SF” and the flow rate value
are indicated if volume
simulation has been
started*
TEST m³h
SF 1 500
Changes every second by with selected
parameter
* Volume pulse simulation is only possible when the
Test Mode
is activated by short-circuiting the contacts SERVICE Flow
simulation is started by pressing and holding the button After its end (in 2 5 minutes), the values of the simulated flow
quantity and energy corresponding to it are recorded
Operating Procedure
Page 26
November 2020
HYB-UM-03193-EN-02