MGS-350 Manual
8
6509-9000 Rev 1
EC Sensors
Formula
Typical Detection Range(s)
Oxygen
O
2
0-30%
Ozone
O
3
0-2 ppm
Phosgene
COCl
2
0-1 ppm
Phosphine
PH
3
0-5 ppm
Sulfur Dioxide
SO
2
0-100 ppm
IR Sensors
Formula
Typical Detection Range(s)
Carbon Dioxide
CO
2
Standard
range:
0-10,000, 0-1% volume
Special
request:
0-5,000ppm, 0-2%, 0-5%,
0-10%
Combustible Gases
Selected
0-100% LEL, 0-100% volume
CAT Sensors
Formula
Typical Detection Range(s)
All Combustible
Gasses
Selected
0-100% LEL
SC Sensors
Example Gases
Typical
Detection
Range(s)
HFCs
R134a, R404A, R407C, R410A, R507
10-10,000 ppm
HCFCs
R22
10-10,000 ppm
CFCs
R11, R12
10-10,000 ppm
Combustibles
Methane (Natural gas), Propane, Butane,
LPG, Isobutane, Ethylene
0-10,000 ppm
Ammonia
NH
3
0-10,000 ppm
Hydrogen
H
2
0-10,000 ppm
VOCs
Acetone, Chloroform, Ethanol, Methanol,
Methyl and Methylene Chloride, Ethyl and
Ethylene Chloride
0-10,000 ppm
NOTE:
The actual factory calibration gas and detection range
is marked on each instrument.
Содержание MGS-350
Страница 10: ...MGS 350 Manual 10 6509 9000 Rev 1 Figure 2 MGS 350 IP66 Housing with Splash Guard...
Страница 11: ...MGS 350 Manual 6509 9000 Rev 1 11 Figure 3 IP66 Splashguard M42 x 1 5 Thread Figure 4 IP66 Duct Mount Housing...
Страница 12: ...MGS 350 Manual 12 6509 9000 Rev 1 Figure 5 Exd Housing Units mm...
Страница 42: ...MGS 350 Manual 42 6509 9000 Rev 1...
Страница 43: ...MGS 350 Manual 6509 9000 Rev 1 43...