Figure 8. Gas Cylinder and Test Hardware
Gas ampoules are convenient and inexpensive alternatives to using gas
cylinders for bump testing.
Step
Bump Testing Using Gas Ampoules
1
Make sure that both the ampoules and the calibration beaker
are clean and dry.
2
Unscrew the beaker hold screw and place the ampoule so
that it sits in the base of the beaker (see Figure 9).
3
Tighten the wing-nut screw onto the ampoule without
breaking it.
4
Remove the enclosure lid of the gas detector (not in an
exhaust area).
5
Connect a voltmeter to monitor sensor response. Monitor 0-
10v on CON3 across pins 1 and 3.
6
Place the beaker over the sensor head using the multi sensor
adaptor to fit the sensor, or, if an Exd, IP66 or Remote sensor
head version, screw the beaker on the remote sensor head
M42 thread or M35 thread adaptor. It should be as tight
fitting as possible to allow maximum gas exposure.