Figure 13. SC Sensor Board
4.6. Calibration: Gas Method for CAT Sensors
There are two adjustments required: the zero and the span. They are
performed on the sensor board but are monitored on the motherboard at
the 0-10V output.
IMPORTANT:
R7 is used to set the supply voltage to the
sensor and can be monitored between TP0 (negative) and
Link 4 (positive). THIS IS FACTORY SET AND SHOULD
NOT BE ADJUSTED.