2.3 Connecting Instrument to Medical Gas Outlet
Before proceeding, turn ON the instrument and allow it warm-up per
Section 2.2. The pump should be OFF at this time.
After warming up, connect the instrument’s gas inlet through the sup-
plied
fl
ow restrictor to the medical gas outlet as follows:
CAUTION:
Never apply gas directly from the medical gas
outlet to the instrument, as internal damage to the instrument
can result. The maximum pressure that can be applied to the
instrument is 2 psig, which corresponds to a
fl
ow of between
100–200 cc/min.
1. Attach the threaded end of the
fl
ow restrictor (supplied with instru-
ment) to a standard medical gas outlet
fi
tting, then attach
fi
tting to
gas outlet on wall.
2. Connect a 2 to 3 foot length of clear,
fl
exible
1
/
8
" tubing (user supplied)
to the barbed end of the
fl
ow restrictor .
3. Attach the other end of the plastic tubing to the instrument’s gas inlet
fi
tting (Figure 1, Item A).
Important!
If the quality of the gas sample is
suspect, protect the instrument by inserting the supplied PTFE particu-
late
fi
lter in-line using appropriate
fi
ttings and tubing.
Figure 2. Connecting Instrument to a Medical Gas Outlet
MGC
2
3
FLOW
RESTRICTOR
2 TO 3 FOOT
SECTION OF
1/8" TUBING
Attach Flow Restrictor to
a standard Medical Gas Outlet
Fitting, then attach fitting to
gas outlet on wall.
4