Hi Flow Sampler
Calibration
Instruction 0055-9017
45
5.
Disconnect the hose from the BACKGROUND inlet port and connect
it to the GAS inlet port.
6.
Again wait several minutes for the gas reading to stabilize. At this
time the gas concentration shown on the display should match the
concentration stamped on the calibration gas cylinder.
1 1 / 2 2 / 0 4 1 0 : 4 5 : 3 0
B a c k ( % ) 0 . 0 0
L e a k ( % ) 2 . 5 0
[ 0 0 0 5 ] ( M )
S - -
7.
Disconnect the hose from the GAS port, and then allow the pumps to
run until both the
Back
and
Leak
readings fall to zero percent.
8.
Press the
ESC
key
three
times to return to the Main Screen.
Note:
If gas is still present in either sensor chamber when
ESC
is pressed,
the instrument will automatically begin to purge the sensors until all gas is
removed from the instrument.
4.4
Gas Calibration Procedure
This procedure calibrates both the leak-gas sensor and the background-
gas sensor at 2.5% and 100% CH
4
. After setting up the calibration
equipment as described in Section 4.2, proceed to calibrate the sensors
as follows:
1.
Before turning ON the instrument, be sure that the calibration
equipment is
not
connected to the instrument’s GAS or
BACKGROUND inlet port, and that the instrument is in an area of
clean air.
2.
Turn ON the instrument and wait its warm-up period to complete
before proceeding to Step 3.
3.
From the Main Screen select
Menu
>
Calibration
>
Calibrate
Sensors
to display the Sensor Calibration Screen.
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
B a c k g n d @ 2 . 5 % C H 4
B a c k g n d @ 1 0 0 % C H 4
L e a k @ 2 . 5 % C H 4
L e a k @ 1 0 0 % C H 4
[ 0 0 0 5 ] ( A )
S - -
4.
Using the
keys, highlight the sensor to be calibrated along with its
gas level. (In the example above, the background sensor will be
calibrated using 2.5% CH
4
.)
Содержание HI FLOW Sampler
Страница 8: ...Introduction Hi Flow Sampler Instruction 0055 9017 4 Notes...
Страница 52: ...Calibration Hi Flow Sampler Instruction 0055 9017 48 Notes...
Страница 62: ...Maintenance Hi Flow Sampler Instruction 0055 9017 58 Figure 5 9 Gas Flow System...
Страница 68: ...Troubleshooting Hi Flow Sampler Instruction 0055 9017 64 Notes...
Страница 71: ......