Operation
Hi Flow Sampler
Instruction 0055-9017
42
The GPS data is saved along with all other measured and calculated
values at the conclusion of an automatic test, or when
Save
is selected
during a manual test.
3.21 Barcode
Scanner
Data
Note:
Barcode scanning can only be performed if
operating in the
Expanded Menu Mode. Refer to Section 3.8.
When a barcode scanner is connected to the instrument’s SCANNER
port, the instrument acknowledges this connection by displaying at the
bottom left corner of the LCD.
When the barcode scanner is swiped across a barcode, the message
“Barcode Scanned” briefly appears at the bottom of the LCD. The barcode
data is stored and will be associated with all following tests until another
barcode is scanned, or until the instrument is turned OFF. The barcode
data is also saved along with all other measured and calculated values at
the conclusion of an automatic test, or when
Save
is selected during a
manual test.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
L e a k ( % ) 0 . 0 0
L e a k ( N o t A v a i l . )
S p e e d ( L O < > H I )
S a v e
S t a r t
B a r c o d e S c a n n e d
[ 0 0 0 1 ] ( A )