Calibration
Instruction 0055-9017
43
4 Calibration
It is recommended that the Hi Flow Sampler be calibrated by your
nearest Bacharach Service Center every 30 days. Calibration, however,
can be performed in the field if your facility has the necessary equipment
and qualified personnel to perform the procedures described in the
following sections.
The following procedures describe how to:
View the dates on which calibrations were performed
Verify calibration
Calibrate the background and leak gas sensors
Display the gain values of the calibration amplifier circuits
Erase calibration
4.1
View Calibration Dates
Do the following to view the dates on which calibrations were performed
on the background-gas and leak-gas sensors at both low (2.5% CH
4
) and
high (100% CH
4
) calibration gas levels:
1.
From the Main Screen select
Menu
>
Calibration
>
Calibration
Dates
.
2.
The following Calibration Dates Screen is displayed where mm/dd/yy
is the month, day, and year calibration was performed:
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
B a c k L o w m m / d d / y y
B a c k H i g h m m / d d / y y
L e a k L o w m m / d d / y y
L e a k H i g h m m / d d / y y
[ 0 0 0 5 ] ( A )
S - -
3.
Press the
ESC
key twice to return to the Main Screen.
4.2
Gas Calibration Equipment Setup
The following equipment is recommended to both check the calibration
and perform a calibration procedure on the Hi Flow Sampler:
Calibration Kit, P/N 0055-8002
Calibration Gas:
2.5% CH
4
in air, P/N 0051-1121
100% CH
4
, P/N00 55-0060
Zero Air, P/N 0051-7131 (optional)
Connect the components of the calibration kit, calibration gas cylinder,
and Hi Flow Sampler together as shown in Figure 4-1. DO NOT, however,
attach the calibration equipment to the instrument until instructed to do
so in the calibration procedure.
Содержание HI FLOW Sampler
Страница 8: ...Introduction Hi Flow Sampler Instruction 0055 9017 4 Notes...
Страница 52: ...Calibration Hi Flow Sampler Instruction 0055 9017 48 Notes...
Страница 62: ...Maintenance Hi Flow Sampler Instruction 0055 9017 58 Figure 5 9 Gas Flow System...
Страница 68: ...Troubleshooting Hi Flow Sampler Instruction 0055 9017 64 Notes...
Страница 71: ......