H25-IR
Operation
Instruction 3015-5286
41
5.1.7 Auto Gain
IMPORTANT! The instrument must be sampling fresh air (zero gas) when the Auto Gain function is selected.
Selecting the Auto Gain function automatically sets the IR Detector’s voltage to its baseline 0 ppm level by
controlling the instrument’s digital potentiometer (DIGIPOT) circuit.
5.2 Factory Menu
The Factory menu is used to set up or troubleshoot the instrument. This menu appears only when activated by
factory personnel or service technicians who know the correct activation procedure.
When activated, display the Factory menu by first pressing
both
the
ENT
and
ESC
buttons on either the
instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the
Up
and
Down
buttons to highlight “FACTORY” and then press
ENT
to display the Factory menu.
Main Menu:
SYSTEM
LOG
SETUPS
CAL
DATA
DIAGNOSTICS
FACTORY
5.2.1 DIGIPOT
IMPORTANT! The instrument must be sampling fresh air (zero gas) when adjusting the DIGIPOT value.
The DIGIPOT function allows
manual
and a more precise adjustment of the digital potentiometer, which
controls the voltage of the IR Detector. Use the
Up
and
Down
buttons to adjust DIGIPOT value until the IR
VOLT reading is 4.200 ±0.0500 volts. Press
ENT
to save the new DIGIPOT value.
5.2.2 Leak Rate
The Leak Rate value is set to match the leak rate of the internal leak source as indicated on the tag attached to
the leak source. Use the
Left
and
Right
buttons to move the blinking cursor across the screen to highlight the
digit to be modified, and then use the
Up
and
Down
buttons to modify that digit. Press
ENT
to save the new
Leak Rate value.
LEAK RATE
2
5.5 g/yr
DIGIPOT=197
IR VOLT=4.2000
Factory Menu:
DIGIPOT
LEAK
RATE
LEAK
GAS
LEAK
TEMP
LEAK
TEMPCO
SENSOR
CAL
MODEL
SELECT
PROBE
TYPE
FIRMWARE
DIGIPOT=197
IR VOLT=4.2000
Содержание H25-IR
Страница 6: ...Contents H25 IR iv Instruction 3015 5286 Notes ...
Страница 26: ...Preparing for Operation H25 IR 20 Instruction 3015 5286 Notes ...
Страница 34: ...Calibration H25 IR 28 Instruction 3015 5286 Notes ...
Страница 43: ...H25 IR Operation Instruction 3015 5286 37 Notes ...
Страница 68: ...Specifications H25 IR 62 Instruction 3015 5286 Notes ...