5.3 Refilling the Tank With Freon
This maintenance operation is applicable only if the instrument has the auto calibration option. The R134a
held in the tank for calibration use is consumed at a rate of 10 x 10-5 std. ml/s. When the calibration gas
secondary pressure indication on the bar graph drops to the 1 00% point ( 98 kPa I 1.0 kgf/CM2 1), refill the
tank with R - 134a by the following procedure. For filling, a dedicated liquid-freon-filling jig (part number :
E708OFZ ) is necessary.
Note: Part number : The item with part number E70BOFZ is an accessory attached to the LS - 20B leak
standard, which contains auxiliaries other than the freon filling jig.
(1)
Remove the instrument cover.
(2)
Fully close the valve on the tank. Turn the handle on the tank top fully clockwise close.
(3)
Close the regulator valve by pulling up on the knob and turning counterclockwise.
(4) Loosen the two tank mounting screws, and remove the tank, being very careful not to lose the 0-ring at
the tank's gas outlet port.
In the next step, you will begin the filling task. Since a large amount of freon gas will leak out during
this process, you should do the work in a location well removed from the leak inspection area.
(5) Open the tank valve (turning the handle) to discharge the remaining R - 134a (gas).
(6) Weigh the tank before filling, so that you will be able to find out how much it has been filled.
(7) Mount the freon filler fitting on a service cylinder containing R - 134a. Before you do this, first make
sure that the filler-fitting valve is closed. Then set the fitting on the cylinder and make sure that the
three hooks on the clamp securely engage the cylinder top cover. Next, turn the valve body clockwise
to pierce through into the cylinder.
(8)
Connect the service cylinder and tank using the coupler on the filler fitting. To do this, first remove the
plug from the tank, and attach the coupler in its place. Screw the coupler on and tighten lightly with a
wrench. Next, Connect the valve that you mounted on the service cylinder to the coupler. Screw them
together by hand.
(9)
Turn the service cylinder upside down, and allow the tank to fill with liquid R - 134a. First, fully open the
valve attached to the service cylinder. Next, very slightly open the valve on the tank (turning the
Содержание H-25C
Страница 14: ...H 25C Instruction Manual Instruction 3015 0216 14 1 6 Internal Flow Paths Measured Gas Internal Flow Path...
Страница 22: ...H 25C Instruction Manual Instruction 3015 0216 22 When power is applied the display will appear as shown below...
Страница 27: ...H 25C Instruction Manual Instruction 3015 0216 27 BACHARACH...
Страница 58: ...H 25C Instruction Manual Instruction 3015 0216 58 Notes...