background image

 
 

 

 

H-25C Instruction Manual

 

Instruction 3015-0216 

31

3.2 Manual 

Calibration 

 

 

Use the manual calibration method ff the instrument is not equipped with the auto calibration option. 

 

Manual calibration is possible in both A-type and B-type measurement modes. 

 

Note : An LS - 20B Leak Standard is required for manual calibration. 

3.2.1 Setting the Switches 

Manual calibration is done in manual calibration mode.  Set the " CAL " switch in the setup area to the " 
MANUAL " position. 

 

Also set the following three switches to match the type of gas, leak flow, and leak flow units of the LS - 
20B to be used for calibration. 

 

  Set the " GAS " switch to the position for the type of gas. 

  Set the "UNIT" switches for the proper units. 

  Set the "ALARM' switches to the leak flow value. 

 

3.2.2 Operator Actions and Calibration Operation 
 

 

 

 

 

Start the calibration procedure as follows: 

 
 

   

 

(1)  Insert the probe into the H-25C "CAL PORT" to apply an ON signal through the spring switch. 

 
 

   

 

Note : When an S-probe is used, carry out this operation after removing the tube for introducing gas. 

 
 

   

 

(2)  After applying the ON signal, immediately pull the probe away from the "CAL PORT" and, with the 

instrument in search mode (probe switch not depressed), let the probe suck in clean air.  The 
instrument will start its calibration operations after about seven seconds. 

 
 

   

When it starts the calibration operation, the instrument will output a tone from the 

 

 

speaker.  At this time, you must perform the following actions: 

 
 

   

(3)  With the probe switch not pressed, touch the probe tip to the LS - 20B PROBE port.  Keep the probe 

touching until the reading on the display stabilizes. 

 
 

   

(4)  When the value stabilizes (wait at least ten seconds), press the probe switch once (for less than 0.5 

sec.). The instrument will store the calibration data at that time. 

 
 

   

(5)  Immediately pull out the probe and allow it to suck in clean air. 

 

After completing the calibration operation, the instrument will go into measurement mode.  If you changed 
any switch settings, return them to their original condition. 

 
 

3.2.3 What to Do If CAL FAIL Appears 

 

If CAL FAIL appears during calibration, calibration is not performed.  When the CAL FAIL indicator 
appears, one of the following three fail codes is displayed on the data display area while the instrument 
stabilizes. 

 

 

E - 1 (no sensor sensitivity) 

 

E - 2 (sensor sensitivity too low) 

 

E - 3 (sensor sensitivity too high) 

 

 

 

Refer to Subsection 3.1.3 for details of what to do when the fail code appears. 

 
 

 

Содержание H-25C

Страница 1: ...H 25C INDUSTRIAL HALOGEN LEAK DETECTOR Instruction 3015 0216 Operation Maintenance Rev 0 Product Leadership z Training z Service z Reliability...

Страница 2: ...6 2 2 4 Setting the RELAY Relay State Selector switch 17 2 2 5 Setting the MODE Measurement Mode Selector switch 18 2 2 6 Setting the Alarm Point 18 2 2 7 Setting the CAL Calibration Mode Selectorl sw...

Страница 3: ...be sucked in 49 6 2 Possible Causes of Malfunctions and What to do If They Occur 50 6 2 1 Does not operate at all 50 6 2 2 Does not detect leaks 50 6 2 3 Measured value is too low 50 51 6 2 4 Measure...

Страница 4: ...is manual these will be hereafter written as R 134a and R 12 2 Unpacking and Initial Checks Your H 25C Halogen Leak Detector was carefully inspected at the factory before shipment and packed so as to...

Страница 5: ...filters on hand at all times Note that the sensor will deteriorate even while in storage if exposed to humidity Always store the sensor in the sealed condition in which you received it 5 Operating Pr...

Страница 6: ...idity To obtain the best leak detection performance keep the following in mind Ventilate the work area well so that contaminants cannot accumulate in the air The surface on which you place the H 25C s...

Страница 7: ...H 25C Instruction Manual Instruction 3015 0216 7 The parts in instruments manufactured before September 94 are different...

Страница 8: ...with contact output Factory setting A type mode 1 Factory setting 4 hours Factory setting 5 times Begin warm up Measurement and Calibration Operator action Data Display Bar Graph Display Speaker Tone...

Страница 9: ...H 25C Instruction Manual Instruction 3015 0216 9 7 Special Symbols and Formatting Used in This Manual The Display State Lit or Flashing is Indicated in Figures as follows...

Страница 10: ...H 25C Instruction Manual Instruction 3015 0216 10 1 NAMES AND FUNCTIONS OF COMPONENTS Model codes specifications gases measured calibration gas flow 1 1 Setup Area Details...

Страница 11: ...ration possible Contact Output N 1 No Relay Yes Dry Contact Relay Output Supply voltage 5 6 7 100 115 V 50 60 Hz 120 240 V 50 60 Hz 110 220 V 50 60 Hz Analog output N 1 No Yes 0 1V DC Analog Output In...

Страница 12: ...plied as option if specified Manual calibration using LS20B Leak Standard 0 to 1V DC output resistance 1 max function supplied as option if specified One Alarm point output generated when measurement...

Страница 13: ...Filter O ring for L type probe 3015 1705 Unit of purchase 1 set one piece each Activated Charcoal pack 3015 2418 Unit of purchase 1 set five pieces Sensor 3015 0807 Unit of purchase 1 piece Fuse rati...

Страница 14: ...H 25C Instruction Manual Instruction 3015 0216 14 1 6 Internal Flow Paths Measured Gas Internal Flow Path...

Страница 15: ...the instrument is shipped with the sensor removed and sealed in a humidity excluding bag Install the sensor as follows 1 Remove the sensor changer cover 2 Remove the protective cap attached to the se...

Страница 16: ...e The GAS switch is at the right of the setup area on the main unit front Open the setup area cover to gain access to it Set the GAS switch lever to the R134a position if using the instrument for R134...

Страница 17: ...set up at the factory as a normally open N contact selecting energize will result in the contacts closing when the signal is output whereas selecting de energize will result in the contacts opening I...

Страница 18: ...n the bar graph The value at which the alarm output trips is called the alarm output trips is called the alarm point Although this point can be freely set by the user there are restrictions on the low...

Страница 19: ...position 2 2 8 Setting the CAL INTVL Calibration Request Interval Switch The H 25C can issue requests for calibration operations at fixed intervals when A type measurement mode is selected At the time...

Страница 20: ...times integer factor the sensitivity of pass fail mode used to measure leak flows However the sensitivity multiplication factor is restricted by the alarm contact setpoint Caution In search mode the...

Страница 21: ...power receptacle 2 5 Applying Power Turning the Power Switch ON Note When using this instrument always keep in mind that it contains parts that operate at high temperatures the sensor and circuits whe...

Страница 22: ...H 25C Instruction Manual Instruction 3015 0216 22 When power is applied the display will appear as shown below...

Страница 23: ...ve is open at the time that you open the valve on the calibration gas tank a pressure of nearly 600KPa 6kgf cm2 may be imposed on the calibration gas tubing system Even though the tubing system will n...

Страница 24: ...RT Therefore once the calibration gas secondary pressure has been adjusted that adjustment need not be done again until the calibration gas R134a is exhausted However when the instrument not adjusted...

Страница 25: ...g the variable resistor labeled AUDIO in the setup area 2 9 Allowing Air to Flow at the End of Preparation Completing the setup operations through Section 2 8 completes the preparations for use If you...

Страница 26: ...again after repeated fixed intervals calibration request interval The calibration request message will also appear if the measured value in search mode is at or above the trigger point see Subsection...

Страница 27: ...H 25C Instruction Manual Instruction 3015 0216 27 BACHARACH...

Страница 28: ...ation gas used in the H 25C s optional built in leak standard is R 134a At a secondary pressure of 170 kPa 1 7 kgf cm2 this cal gas will flow out from the CAL PORT at a rate of 9 X 10 5 to 11 x 10 5 s...

Страница 29: ...he countdown starting from 7 on the display reaches 0 Leave the instrument in search mode do not press the pass fail switch on the probe When the countdown reaches 0 the instrument s internal circuits...

Страница 30: ...s displayed The capillary is plugged and the calibration gas cannot flow out from the CAL PORT 2 E 2 Sensor Sensitivity Too Low or E 3 Sensor Sensitivity Too High This fail code is displayed if the se...

Страница 31: ...L PORT and with the instrument in search mode probe switch not depressed let the probe suck in clean air The instrument will start its calibration operations after about seven seconds When it starts t...

Страница 32: ...on approaches the allowable limit an error code A 1 to A 3 is displayed for about four seconds before completing calibration The two alarm codes are as follows take appropriate action depending on the...

Страница 33: ...d the atmosphere of unwanted gases where leakage inspections are made are given at the end of the manual after chapter 6 4 1 2 Getting Ready for Measurement 1 Connecting the Power Cable Supply the H 2...

Страница 34: ...go into measurement mode after about 180 seconds of warm up During the warm up time or immediately after transferred to the measurement mode check the following Make Sure that the Fan is Running The f...

Страница 35: ...e If the intake flow remains below about one quarter of this value for about 30 seconds AIR FLOW will be displayed If AIR FLOW is displayed replace the probe filter If the intake flow does not recover...

Страница 36: ...the probe indicator lamp lights The tone and the lamp are switched off when the value is 20 below the alarm point If the H 25C is equipped with an alarm contact output then if the leak rate reaches t...

Страница 37: ...h the tone change If you have difficulty in searching for a leak due to unwanted gas in the atmosphere I Ambient gas levels high enough to exceed the search trigger level will not only interfere with...

Страница 38: ...herefore you must exercise care in the use of this feature if it is important to know for sure whether any leak is present Calibration Request Operation functions in A type mode A calibration request...

Страница 39: ...ate is the same as in B type mode Operation in pass fail mode is as follows In pass fail mode the 100 point on the bar graph display becomes the alarm point The measured value in the display area is d...

Страница 40: ...display section set the UNIT switch as shown in the figure below Note that the displayed value is the result of computation based on the secondary pressure of the calibration gas The displayed value m...

Страница 41: ...s deteriorated See Section 5 4 for corrective action 4 2 7 Turning the Power Switch OFF The usable life of the sensor and the activated charcoal filter are highly dependent on the time that power is O...

Страница 42: ...spring switch In case of the S type probe the tube for introducing gas must be removed to perform this operation 2 After providing ON signal immediately remove the probe from the CAL PORT and suck in...

Страница 43: ...in by the air is removed by the probe filter When the material removed begins to clog the filter sufficiently that a condition in which the flow drops to around 1 4 that rate continues longer than 30...

Страница 44: ...its effectiveness at conversion In general if you are using the H 25C seven hours a day an activated charcoal filter can be used for 30 to 60 days We recommend that you schedule periodic replacements...

Страница 45: ...then push them down into the case see figure 5 4 c Push down the whole filter until it is lower than the top edge of the case d Replace the cover so the open end of the tube beneath the cover touches...

Страница 46: ...t step you will begin the filling task Since a large amount of freon gas will leak out during this process you should do the work in a location well removed from the leak inspection area 5 Open the ta...

Страница 47: ...Note that proper calibration will be impossible if any air is left in the tubing After adjustment loosen the gas bleed plug for about three seconds repeat s action two or three time 5 4 Replacing the...

Страница 48: ...d in dry the inner tube before replacing the filter Note the following when replacing the filter Turn the power switch off Since the disc filter distinguishes the inflow direction from the outflow dir...

Страница 49: ...ely divide the measured gas flow If you allow liquid to be sucked into the instrument contaminants will be deposited on the manifold surfaces changing the division ratio so that accurate measurement w...

Страница 50: ...calibrate to proper value See Subsection 6 2 10 6 2 1 Does Not Operate at All The three most likely causes are 1 Blown Fuse After making sure that power is actually being supplied check the fuse See S...

Страница 51: ...working or performing poorly Replace with good pump Note No adjustment required after replacement 4 Manifold internal flow path clogged Return to Bacharach for repair component replacement 5 Flow sens...

Страница 52: ...Used Up Too Quickly The most likely two causes are 1 Leakage from gas bleed plug at capillary receiver section Check first whether plug is loose If not replace 0 ring 2 Leakage from tank seal gas outl...

Страница 53: ...ed it is recommended that the facilities described below be provided Example 1 of Proper Facility Reference Figure 1 shows an example of a facility that helps extend the service life of sensors and ac...

Страница 54: ...rolong the service life of sensors and activated charcoal filters If the facility to be provided is similar to that shown in Reference Figure 2 take the following requirements into consideration 1 The...

Страница 55: ...15 2560 1 1 1 2 1 1 1 1 1 1 1 1 Pump Assembly Regulator Assembly Regulator O Ring Flow Sensor Assembly Fun Assembly Activated charcoal Filter Sheet filter Activated charcoal Pack 5 pcs Fuse 3 15 A 250...

Страница 56: ...5 0947 3015 1545 3015 2096 3015 2102 3015 2108 3015 2114 3015 1194 1 1 1 1 1 1 1 1 Tank Assembly w freon O Ring O Ring O Ring O Ring Printed circuit Assembly Printed circuit Assembly For 100 115V powe...

Страница 57: ...B 4 Houston TX 77092 Tel 713 683 8141 Fax 713 683 9437 E mail TXSERVICE bacharach inc com CANADA Bacharach of Canada Sales Service Center 250 Shields Court Unit 3 Markham Ontario L3R 9W7 CANADA Tel 90...

Страница 58: ...H 25C Instruction Manual Instruction 3015 0216 58 Notes...

Страница 59: ...World Headquarters 621 Hunt Valley Circle New Kensington PA 15068 Ph 724 334 5000 Fax 724 334 5001 Toll Free 1 800 736 4666 Website www bacharach inc com E mail help bacharach inc com Printed in U S A...

Отзывы: