" Calibration requests " are being issued too frequently. Activated charcoal filter performance is
deteriorating.
•
" CAL FAIL " indicator is displayed (flashing).
Calibration gas freon in the tank has been used up. (See Section 5.3 for corrective action.)
•
Unable to calibrate ("CAL FAIL" displayed).
Sensor performance has deteriorated. (See Section 5.4 for corrective action.)
4.2.7
Turning the Power Switch OFF
The usable life of the sensor and the activated charcoal filter are highly dependent on the time that
power is ON (the time that gases to which the instrument is sensitive are being drawn in). We
therefore recommend that you turn the instrument power switch OFF if you do not intend to make any
measurements for a period of several hours.
Note that before turning the power switch OFF, you should always let the instrument draw in clean air
for at least one minute.