6 Tangens Delta
46
DHV0053 b2 EN
©
b2electronic
gmbh
Currently Measuring…… Phase Selection
The phase that is currently being tested by the HVA28TD needs to be selected.
All new measurement data will be put into the corresponding report phase
section (tab).
If the operator changes the phase during an active HVA28TD measurement, the
change will only occur when the current HVA28TD voltage block is finished to
avoid splitting up of the respective HVA28TD measurement data block. The
block “under the wrong” phase tab can then be moved to the correct phase
section using the “Change Phase” button in the block header as a later stage.
Graphical View / Diagram
This screen provides a graphical diagram of the Mean HVA28TD measurement
data versus the applied test voltage.
The values of all three phases can be viewed and compared. The graph
automatically scales and draws the mean HVA28TD values for each voltage
block.
Phase Selection for Diagram View
Toggle ON/OFF the various phases to display / block out the relevant phase
information.
Copy Graph to Clipboard
Use this function to make a copy the currently shown HVA28TD diagram into
the Windows Clipboard. It can then be easily put into many other applications
(like Word, Excel) by selecting Paste in the target program.