PGE-W series crane scale Technical Manual
- 9 -
F7 ci
Set new password
P1---
000
F8 cl
Calibrate,
Unld
calibrate weight
load
complete
See detail in Calibration
9. B. CALIBRATION
1) Turn
ON
the Scale.
2) Press
ZERO
and
CHANGE
keys both during self checking.
3)
Display shows “P1---”, waiting for password.
4)
Enter the password, use
UNIT
key to move the active digit, use
CHANGE
key to increment a digit, use
ZERO
key to sure, default password is “000”
5) Display will shows “F0 di”, press
CHANGE
key until display shows “F8 Cl”
6) Press
ZERO
key to enter calibrate, display shows “Unld”
7)
Empty the hook, after
STABLE
indicator on, press
ZERO
key to sure, after a
few seconds, zero calibration completely.
8)
After zero calibration, display will show span calibration weight value. You
can change this weight value, (use
UNIT
key to move the active digit, use
CHANGE
key to increment a digit), enter the calibration weight value. Then
press
ZERO
key to sure, display show “LOAD”, now to add calibration weight on
the hook. When the stable indicator is on press
ZERO
key again. Span
calibration will complete with in few seconds.
9. B. CHANGE CAPACITY
1) Open the Top cover.
2) Short J2 Pin
3) Then turn on the power, press the
CHANGE
and
ZERO
key both when
display start self checking. Display shows “
P3---
”, waiting for input password.
4) Input password, use
UNIT
key to move the active digit, use
CHANG
E key to
increment a digit, use
ZERO
key to sure, default password is
“
999
”
5) Display will shows the current capacity, you can use
CHANGE
key to change
capacity, use
UNIT
key to move decimal point, after you select, press
Zero
key to sure.
6) Then scale will enter parameter setting mode Open J2 jumper