Technical Safety Check TSC
Infusomat® fmS, 2.3 gb
7 - 3
Sheet 3 of 4
M670 00 00 92 F04 C 3891 2189 (GB)
Index d
(Master – to be added to the documentation)
Delivery accuracy
Ambient temperature 20 ... 25 °C
Delivery rate: 250 ml/h
Measured volume: 25 ml
❒
Divergence (± 5%)
_______ %
Drop sensor
Delivery rate: 400 ml/h
❒
Drop sensor alarm, occlusion ______ s
(< 5s)
❒
Drop sensor alarm, flow
Air inline sensor
Delivery rate: 250 ml/h
❒
Water value (> 481 mV) _______ mV
❒
Air alarm
❒
Air value (< 78 mV)
_______ mV
❒
Threshold value (< 182 mV) _____ mV
Battery test
Operation mains/battery/mains:
❒
Uninterrupted operation
Switch on unit without mains connection
❒
Self-test is executed
Visual Inspection
Electrical Safety
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Functional Inspection
(Part 3 of 3)
Содержание Infusomat fmS
Страница 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Страница 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Страница 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Страница 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Страница 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Страница 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Страница 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Страница 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Страница 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Страница 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...