Unit Elements
4
Infusomat® fmS 2.3 gb
4 - 13
-
Air alarm: air bubbles in ml and air rate in ml/h respectively
-
User language
-
Special functions (ON/OFF)
-
Menu
-
Staff call type
Note
If data is not entered, ”Calibration data faulty” may be displayed
after the unit is switched on again.
9.
After ending the Service Program save the data.
The counters for operating hours, battery operation and pump
head cycles are reset to zero when the pump unit is exchanged. If
”Calibration data faulty” is displayed, select the Service Program
and quit with ”Save? Yes”.
Check
Safety check, pump unit check.