12
en
3.3
Single-use products
►
Do not reuse the product.
The reprocessing of the product affects its functionality. Risk of injury, ill-
ness or death due to soiling and/or impaired functionality of the product.
►
Do not reprocess the product.
3.4
Reusable products
Influences of the reprocessing which lead to damage to the product are
not known.
A careful visual and functional inspection before the next use is the best
opportunity to recognize a product that is no longer functional, see
Inspection.
3.5
Preparations at the place of use
►
If applicable, rinse non-visible surfaces preferably with deionized
water, with a disposable syringe for example.
►
Remove any visible surgical residues to the extent possible with a
damp, lint-free cloth.
►
Transport the dry product in a sealed waste container for cleaning and
disinfection within 6 hours.
3.6
Preparing for cleaning
►
Disassemble the product prior to cleaning, see Disassembly.
►
Disassemble the product prior to cleaning.
►
Open up product with hinges.
►
Open all valves/faucets.
3.7
Disassembly
3.7.1
S
4
C navigated tapper
Ø
3.5 mm FW655R
►
Loosen nut
19
and unscrew from the sleeve
17
.
►
Remove sleeve
17
in the direction of the arrow.
Note
Spring
18
is fixed on shaft.
Fig. 14 Disassembling the S
4
C tapper (Tap)
3.7.2
S
4
C navigated screwdriver FW656R
►
Retract and hold green holding sleeve
22
against the spring pressure.
►
Unscrew screw sleeve
20
at the working end of the S
4
C screwdriver
and remove from the shaft.
►
Release the green holding sleeve
22
.
►
Push the green holding sleeve
22
with the retaining tongues in the
direction of the instrument's working end and remove from the screw-
driver shaft.
Fig. 15 Disassembling the S
4
C screwdriver
3.7.3
S
4
C navigated drill sleeve
Ø
3.5 mm (FW664R)/
Ø
4.0 mm
(FW665R)
►
Turn guide sleeve
24
clockwise and remove. Be aware that it is a left-
handed thread.
►
Loosen knob
25
counterclockwise, unscrew and remove from the han-
dle by pulling in the direction of the arrow.
Fig. 16 Disassembling the S
4
C drill guide
3.7.4
S
4
C navigated guide sleeve for smooth-shaft screws, FW658R
►
Turn inner sleeve
26
to the "remove" position and remove from the
outer sleeve in the direction of the arrow.
►
Loosen knob
25
counterclockwise, unscrew and remove from the han-
dle by pulling in the direction of the arrow.
Fig. 17 Disassembling the guide sleeve for smoothshank screws
Art. no.
Designation
FW086SU
Drill
Ø
2.9 mm for smooth-shaft screws
FW088SU
Favored Angle screw drill
Ø
2.9 mm for
Ø
4 mm screws
FW662SU
S
4
C1/C2 drill
Ø
2.4 mm
FW666SU
Drill,
Ø
2.4 mm
FW667SU
Drill,
Ø
2.9 mm
19
18
17
A
20
21
22
23
D
25
24
F/G
25
26
I
Содержание Aesculap S4 Cervical
Страница 2: ...A B C D E F H K J G I ...
Страница 3: ...L M N O P Q R S ...