Imporrtant Safety Infomation
•
The lasers inside can cause eye damage before you have a chance to blink!
•
There are high voltages inside.
•
There are no user serviceable parts inside.
•
The unit will be damaged if you attempt to open the protective housing.
Interlock Failure Symptoms
If you experience any of the following symptoms, you may have an interlock failure. It is unsafe to continue to use
the GenePix 4200A with only one interlock functioning. Please contact Technical Support immediately at either of
the numbers indicated on the rear panel of the GenePix 4200A.
1.
Mirror and lens assembly keeps moving below the slide after the door is opened.
2.
The blue “Scanning” LED at the front of the GenePix 4200A doesn’t extinguish when the door is open.
3.
You cannot scan any slides.
4.
You can no longer hear the distinctive metal-on-metal sound that the mechanical interlock makes when it falls
into place. This normally happens when the slide loading door has opened about ¼”.
Lifting Hazard
Warning:
The GenePix Professional 4200 scanner weighs approximately 99 lbs. (45 kg) and requires two persons
to lift. There are four molded handholds on the underside of the scanner (see figure). Use appropriate lifting
technique to avoid injury.
CLASS 1 LASER
PRODUCT
Lasers
Cautions
•
Do not stick anything into the openings or vents of the GenePix 4200A, other than loading the slide.
You may cause an interlock to fail.
•
Do not attempt to move any parts below the slideholder. You may cause an interlock to fail.
GenePix 4200A User’s Guide, Copyright 2005 Axon Instruments / Molecular Devices Corp.