USER MANUAL
31
_________________________________________________________________________________________________________________
Internal calibration options:
Internal calibration of the balance is performed automatically every time the balance is switched on,
additionally after given time interval during work and after every temperature change of more than given
value.
In order to perform internal calibration in any moment, empty the pan and press key twice (one more
pressing terminates calibration).
Press
MENU
key to display function menu and
choose
CALIb
function by pressing
T
key
when it is displayed.
The following options will appear:
-
CAL on
–
perform calibration with external weight
-
CAL Prn
–
printout of calibration report
-
CAL tM
–
set time interval for internal calibration
(1h
–
6h)
-
CAL
C
- set temperature difference for internal
calibration (1
C - 4
C)
-
out
–
switch internal calibration off
for internal calibration
Press
T
key when
CAL tM
option is displayed
.
Predefined time intervals for internal calibration will
be displayed. Select required value pressing
T
key
.
Accordingly choose
CAL
C
option pressing
T
key and selecting values of temperature difference.
Select
out
option to finish.