Buzzer
option is used for selection of
acoustic signalling method. Selection of
Stable OK
option means acoustic signal
activation after indications stabilisation
in range signalled as
OK.
It is possible
to activate the signal immediately after
exceeding the threshold, or total signal
deactivation.
To start working with thresholds
signalling, move cursor to
Activate
option and press
ENTER
.
After each loading, the result of
comparing
with
thresholds
is
signalled.
To finish the function operation,
press
MENU
to enter the user
functions menu, select
Threshold
signalling
function,
and
select
Deactivation
option.
If the balance is equipped with
THRESHOLDS
control connection, the comparison
result may be used to control the external optical signalling device, or other
external devices.
On the outputs
P1
and
P2
the short-circuit states appear, which depend on
comparison results of balance indications with threshold values. The available
operation modes are shown on the states chart.
USER FUNCTIONS \ TRESHOLD SIGNAL.
1. Deactivate
2. Zero treshold : 5g
3. Min treshold : 90g
4. Max treshold : 110g
5. Outputs mode : signalling
6. Buzzer
: stable OK
7. LCD flashing : OFF
8. Exit
Max ... Min ... e= ... d= ..
100.000
g
0%
100%
USER FUNCTIONS \ TRESHOLD SIGNAL.
1. Activate
2. Zero treshold : 5g
3. Min treshold : 90g
4. Max treshold : 110g
5. Outputs mode : signalling
6. Buzzer
: <stable OK>
7. LCD flashing : OFF
8. Exit
USER FUNCTIONS \ TRESHOLD SIGNAL.
1. Activate
2. Zero treshold : 5g
3. Min treshold : 90g
4. Max treshold : 110g
5. Outputs mode : signalling
6. Buzzer
: stable OK
7. LCD flashing : OFF
8. Exit