14.1 Autotaring function
This function automatically keeps zero indication when a pan is empty or zero
indication was forced with
T
key.
Press
Menu
key to enter the user
function menu, choose
Autotaring
and
press
ENTER
key.
Choose
Correction range
using
∨
and
∧
keys
and press
ENTER
key.
Enter maximum zero flow to be
automatically corrected by
Autotaring
function. Allowed range is 0.5 ÷ 5 (in
balance reading units per second).
Choose
Activate
option with a cursor
and press
ENTER.
Indications that are near to zero will
be automatically corrected and zero
indication will be maintained despite
changing environmental conditions
(temperature, air humidity, etc.)
To stop the function press
Menu
key to enter user functions menu, choose
Autotaring
function and then choose
Deactivate
option.
USER FUNCTIONS&OPT.
Autotaring
PCS counting
. . .
USER FUNCTIONS&OPT. \ AUTOTARING
1. Activate
2. Correction range: 2.0 d/sec
3. Exit
USER FUNCTIONS&OPT. \ AUTOTARING
1. Activate
2. Correction range: 2.0 d/sec
3. Exit
Max ... Min ... e= ... d= ..
AUT
0.000
g
0%
100%