X. Error Codes
1. OUTZ
- Zero point has drifted out of the acceptable
range. Try to perform a calibration adjustment. If the
problem persists, the load cell may be permanently
damaged due to overloading.
2. O-Ld
- Overload. The maximum capacity has
been exceeded. Remove the extra weight to avoid
permanently damaging the load cell.
3. Lo
- Batteries are low. Replace the batteries.
XI.
Specifications
MB-100
MB-650
Max. Capacity
100g
650g
Readability
0.01g
0.1g
Min. Weight
0.03g
0.2g
Power
One “CR2032” Lithium Battery
Dimensions
1.4 x 2.9 x 0.5 in.
Platform Dimensions
1.4 x 1.8 in.
Warranty
10 Year Limited Warranty
XII. Capacity/Readability Charts
MB-100
MB-650
100g x 0.01g
3.527oz x 0.001oz
3.215ozt x 0.001ozt
64.30dwt x 0.01dwt
500ct x 0.05ct
1543.2gn x 0.2gn
650g x 0.1g
22.92oz x 0.01oz
20.89ozt x 0.01ozt
418.0dwt x 0.1dwt
3250ct x 0.5ct
10031gn x 2gn