Matchbox Manual

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

MB-100

0.02g

100g

OIML M2

MB-650

0.2g

500g

OIML M2

VIII.  Adjustment

1.  With the scale ON, press and hold the 

ON/OFF 

button to enter calibration mode.

2. 

The LCD will flash “CAL” followed by the required 

calibration weight (see chart above).

3.  Place the required weight on the center of the scale 

platform. The display will flash “F” then return to 

normal weighing. Adjustment is complete! 

NOTE:

 Adjustment using the incorrect mass will 

result in inaccurate readings. Only use listed weight.

IX. Troubleshooting

• 

 

Problem:

 Scale will not turn on

• 

 

Solution: 

Check the batteries and battery contacts.

• 

 

Problem: 

Displayed weight fluctuates randomly

• 

 

Solution: 

Make sure your workspace is stable and    

 free of any air currents or vibration. In some cases, 

nearby radio wave emitting devices can cause 

fluctuations.

• 

 

Problem:

 Displayed weight is inaccurate

• 

 

Solution:

 Perform a calibration adjustment. If 

you are still having trouble, contact our warranty 

department. 

Содержание MB-100

Страница 1: ...American Weigh Scales MB 100 100g x 0 01g MB 650 650g x 0 1g User Manual Matchbox Scale Copyright 2011 American Weigh Scales Inc All rights reserved Rev 3 0 ...

Страница 2: ... for 30 60 seconds before performing calibration if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction Avoid using your scale in close proximity to strong radio wave emitting devices such as cordless phones Always weigh on a flat and level surface free from vibrations and drafts The corn...

Страница 3: ...cription 1 ON OFF Press to turn the scale on or off Hold while scale is on to enter Calibration Mode expert 2 TARE MODE Press briefly to re zero the scale Hold to change weighing unit III Battery Installation Proper Battery Installation Positive Side Facing Out 2 1 g ON OFF TARE MODE ...

Страница 4: ...n a flat level surface and press the ON OFF key to turn the scale on 2 Once the start up process completes and the scale goes to zero you are ready to weigh 3 Place the quantity to be weighed onto the weighing platform 4 Wait a couple seconds for the display to stabilize before taking a weight reading 5 To turn the scale OFF press the ON OFF key again V Changing the Unit of Measure 1 Press and hol...

Страница 5: ...eight of the contents 4 To return to normal weight display gross weight remove any items from the weighing platform and press the TARE MODE key again to re zero the scale VII Calibration Initial calibration is performed at the factory However the calibration can drift slightly due to stresses from shipping Also due to changes in local acceleration of gravity around the world adjustment is recommen...

Страница 6: ...sh F then return to normal weighing Adjustment is complete NOTE Adjustment using the incorrect mass will result in inaccurate readings Only use listed weight IX Troubleshooting Problem Scale will not turn on Solution Check the batteries and battery contacts Problem Displayed weight fluctuates randomly Solution Make sure your workspace is stable and free of any air currents or vibration In some cas...

Страница 7: ...l 3 Lo Batteries are low Replace the batteries XI Specifications MB 100 MB 650 Max Capacity 100g 650g Readability 0 01g 0 1g Min Weight 0 03g 0 2g Power One CR2032 Lithium Battery Dimensions 1 4 x 2 9 x 0 5 in Platform Dimensions 1 4 x 1 8 in Warranty 10 Year Limited Warranty XII Capacity Readability Charts MB 100 MB 650 100g x 0 01g 3 527oz x 0 001oz 3 215ozt x 0 001ozt 64 30dwt x 0 01dwt 500ct x...

Страница 8: ...Matchbox Manual American Weigh Scales Inc Norcross GA 30092 Made In China to our exact specifications and quality control ...

Отзывы: