FRONT PANEL CONTROLS (UNITS WITH -FPD OPTION)
1. POWER Switch
. This is the main power switch. When turning the instrument on,
there is normally a delay of 10 seconds before anything is shown on the main
display, as the internal operating system boots up.
If the main menu does not appear after 30 seconds, turn off the instrument and
leave it off for at least 60 seconds before applying power again.
2. OVERLOAD Indicator
. When the instrument is powered, this indicator is normally
green, indicating normal operation. If this indicator is yellow, an internal automatic
overload protection circuit has been tripped. If the unit is overloaded (by operating at
an exceedingly high duty cycle or by operating into a very low impedance), the
protective circuit will disable the output of the instrument and turn the indicator light
yellow. The light will stay yellow (i.e. output disabled) for about 5 seconds after
which the instrument will attempt to re-enable the output (i.e. light green) for about 1
second. If the overload condition persists, the output will be disabled again (i.e. light
yellow) for another 5 seconds. If the overload condition has been removed, the
instrument will resume normal operation.
24
2
3
4
5
1
8
6
7
9
Содержание AVRQ-4-B
Страница 41: ...41...
Страница 51: ...Bottom side 51...
Страница 67: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY...
Страница 68: ...PCB 104G KEYPAD DISPLAY BOARD...
Страница 69: ...PCB 228C OP AMPS AHV AND XHV UNITS...
Страница 71: ...DUT WIRING ON STANDARD DAUGHTERBOARD PCB 267D...
Страница 72: ...DUT WIRING ON SO 8 DAUGHTERBOARD PCB 314A...
Страница 73: ...DUT WIRING ON WDIP16 DAUGHTERBOARD PCB 325A...
Страница 75: ...PERFORMANCE CHECK SHEET 75...