Page 17
4.2. Supervisor Menu
In the Supervisor Menu there are various submenus available to configure system parameters, specific
sections of the operating modes and the calibration feature (see also chapter 6. Calibration).
4.2.1. Supervisor Menu Levels [LF]
Navigate the options by pressing
ZERO
(◄),
TARE
(►) and increase a numeric value by
SELECT
(▲)
key. Press
UNITS
(
ESC
) to exit the option or the menu and
(
↵
) key
to confirm.
4.2.2. Supervisor Functions Descriptions [LF]
There are 8 functions.
LF 1
1
st
Calibration
:
see chapter 6. Calibration at page 21.
LF 2
Configuration
: use this item to configure system parameters, specific sections of the
operating modes.
See chapter 5. Scale Configuration at page 20.
LF 3
Linearization
: use this to add 1 linearization point for the scale.
CAUTION!
Perform the linearity procedure only if test weights applied to the scale
between the zero and span calibration points are showing slight inaccuracies, such as ±
a few divisions. If large inaccuracies are recorded, this indicates a possible mechanical
problem or possible load cell failure which linearity calibration may not be able to correct.
LF 4
ADC Update Rate
: use this to select the sampling frequency: 15 (1), 30 (2) or 7.5Hz
(3)