4.1.2. User Functions Descriptions [UF]
There are 9 Setup Functions.
UF-1
Diagnostic:
use this menu to check or verify the performance of the indicator. The
diagnostic tests available include: Scale A to D to view output from the connected scale
base or load device,
Internal D-Value
, and
Battery Voltage
.
UF-2
Checkweighing
: the acceptable weight is any weight which falls between the upper and
lower limits. Enter this option to set the items relating to checkweighing:
•
LOW
/
HIGH
limits. Valid beeper modes:
o 000
;
o 001
;
o 002
.
See chapter 3.4. Checkweighing Mode at page 13.
UF-3
Auto Power-Off
: use this to set the length of time before the scale automatically turns
off if not being used. Values can be set between 01 and 99 minutes.
Factory default: AoFF 10
(
10min
)
.
Auto Power-OFF disabled when
AoFF 00
UF-4
Backlight
: use this to set the light mode:
•
OFF
/
Automatic
/
ON
Factory default: Lit A (automatic)
UF-5
Not supported.
UF-6
Not supported.
UF-7
ADC Update Rate
: use this to select the sampling frequency: 15 (1), 30 (2) or 7.5Hz
(3)
See LF 4, Supervisor Functions Descriptions
UF-8
Zero Tracking
: use this to define a ±0 /5div range around zero. When scale weight is
not at the center of zero but inside this range, ½ of the weight will be subtracted until the
weight is inside the center of zero region.
See LF 5, Supervisor Functions Descriptions.
UF-9
Gravitational value
:
use this item to key in a gravity constant value.
CAUTION!
This item is locked when the calibration switch is at position OFF.
See LF 7, Supervisor Functions Descriptions and chapter 6. Calibration on page 21.